DIGITIZATION OF ELECTROCARDIOGRAMS BY DESK-TOP OPTICAL SCANNER

被引:13
作者
WIDMAN, LE [1 ]
HINES, LS [1 ]
机构
[1] VET ADM MED CTR, CARDIOL SERV, ALBUQUERQUE, NM 87108 USA
关键词
DIGITAL SIGNAL PROCESSING; OPTICAL SCANNER; ELECTROCARDIOGRAM; ANALOG-TO-DIGITAL CONVERSION; MICROCOMPUTER; MEDICAL RECORDS;
D O I
10.1016/0022-0736(91)90015-E
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The fidelity of a semiautomated technique for converting paper electrocardiogram (ECG) tracings to digital form by optical scanning was examined. Sample tracings from one nonmechanical and three mechanical ECG writers (recorders) were used. The optically scanned signals were compared with the digitized version (402 Hz, 12-bit precision) of the original analog signals using time- and frequency-domain correlation coefficients and root mean square error. A total of 261 QRS complexes and 207 RR intervals were examined in 21 leads acquired from 8 patients. When data were low-pass filtered at 25 Hz, the correlation coefficients for the 261 QRS complexes were 0.997 +/- 0.005 (mean +/- SD) for the time domain data, 0.992 +/- 0.010 for the complex frequency domain (amplitude and phase) data, and 0.998 +/- 0.002 for the power spectrum. The corresponding correlations for the 207 RR intervals were 0.993 +/- 0.008, 0.992 +/- 0.008, and 0.993 +/- 0.009. The RMS errors, normalized for signal amplitude, were 2.62 +/- 1.28 (percent +/- SD) for QRS complexes and 1.82 +/- 0.87 for RR intervals. The correlations for the mechanical ECG recorder tracings were the same or better than those of the nonmechanical recorder, and the RMS errors were generally smaller. When data were low-pass filtered at 105 Hz, the correlation coefficients ranged from 0.984 to 0.996 for the QRS complexes and 0.982 to 0.988 for RR intervals. Root mean square errors were 4.54 +/- 2.03 and 2.38 +/- 1.14, respectively. For purposes of arrhythmia analysis by QRS classification, digitization of ECG signals by optical scanning appears equivalent to acquisition via standard analog-to-digital conversion.
引用
收藏
页码:325 / 338
页数:14
相关论文
共 50 条
  • [41] Magnetic spring characteristic of an oscillatory actuator with silicone rubber torsion springs for optical scanner applications
    Bu, Yinggang
    Kinjo, Hideyuki
    Oyaizu, Kazuaki
    Inoue, Kaname
    Mizuno, Tsutom
    Duan, Zhihui
    INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2014, 46 (03) : 463 - 470
  • [42] Magnetic actuation of bending and torsional vibrations for 2D optical-scanner application
    Garnier, A
    Bourouina, T
    Fujita, H
    Hiramoto, T
    Orsier, E
    Peuzin, JC
    SENSORS AND ACTUATORS A-PHYSICAL, 2000, 84 (1-2) : 156 - 160
  • [43] Development of a 3D optical scanner for evaluating patient-specific dose distributions
    Chang, Kyung Hwan
    Lee, Suk
    Jung, Hong
    Choo, Yeon-Wook
    Cao, Yuan Jie
    Shim, Jang Bo
    Kim, Kwang Hyeon
    Lee, Nam Kwon
    Park, Young Je
    Kim, Chul Yong
    Cho, Sam Ju
    Lee, Sang Hoon
    Min, Chul Kee
    Kim, Woo Chul
    Cho, Kwang Hwan
    Huh, Hyun Do
    Lim, Sangwook
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (05): : 553 - 559
  • [44] A surface micromachined optical scanner array using photoresist lenses fabricated by a thermal reflow process
    Toshiyoshi, H
    Su, GDJ
    LaCosse, J
    Wu, MC
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2003, 21 (07) : 1700 - 1708
  • [45] Novel Optical Scanner Using Photodiodes Array for Two-Dimensional Measurement of Light Flux Distribution
    Chong, K. K.
    Yew, T. K.
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (08) : 2918 - 2925
  • [46] Study of highly sensitive smell sensing system using gas detector tube combined with optical scanner
    Tanaka, Y
    Nakamoto, T
    Moriizumi, T
    Transducers '05, Digest of Technical Papers, Vols 1 and 2, 2005, : 1880 - 1883
  • [47] An equivalent circuit model for vertical comb Drive MEMS optical scanner controlled by pulse width modulation
    Institute of Industrial Science, University of Tokyo, 4-6-1, Komaba, Meguro-ku, Tokyo 153-8505, Japan
    不详
    不详
    Maruyama, S., 2012, Institute of Electrical Engineers of Japan (132) : 1 - 9
  • [48] A Behavioral Model for Optically Powered OCT Endoscope with a Micro Electrostatic Vertical-Comb Optical Scanner
    Nakada, Muneki
    Chong, Changho
    Morosawa, Atsushi
    Isamoto, Keiji
    Suzuki, Takuya
    Fujita, Hiroyuki
    Toshiyoshi, Hiroshi
    IEEJ TRANSACTIONS ON ELECTRICAL AND ELECTRONIC ENGINEERING, 2014, 9 (04) : 448 - 458
  • [49] A novel optical scanner with integrated two-dimensional magnetostrictive actuation and two-dimensional piezoresistive detection
    Bourouina, T
    Lebrasseur, E
    Reyne, G
    Fujita, H
    Masuzawa, T
    Ludwig, A
    Quandt, E
    Muro, H
    Oki, T
    Asaoka, A
    TRANSDUCERS '01: EUROSENSORS XV, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2001, : 1328 - 1331
  • [50] Accuracy of high-density EEG electrode position measurement using an optical scanner compared with the photogrammetry method
    Gyorfi, Orsolya
    Ip, Cheng-Teng
    Justesen, Anders Bach
    Gam-Jensen, Maria Louise
    Romer, Connie
    Fabricius, Martin
    Pinborg, Lars H.
    Beniczky, Sandor
    CLINICAL NEUROPHYSIOLOGY PRACTICE, 2022, 7 : 135 - 138