THE CLOSING VELOCITY OF MECHANICAL HEART-VALVE LEAFLETS

被引:3
作者
GUO, GX
CHIANG, TH
QUIJANO, RC
HWANG, NHC
机构
[1] UNIV MIAMI,DEPT BIOMED ENGN,CORAL GABLES,FL 33124
[2] BAXTER HEALTHCARE CORP,EDWARDS CARDIOVASC SURG DIV,IRVINE,CA 92714
关键词
MECHANICAL HEART VALVE; LEAFLET CLOSING VELOCITY; MOCK FLOW LOOP TESTING; LASER SWEEPING TECHNIQUE; VALVE HOUSING COMPLIANCE;
D O I
10.1016/1350-4533(94)90069-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The closing motion of the occluder leaflets in bileaflet type mechanical heart valves (MHV) runs monitored with a laser sweeping technique. The angular displacements of the leaflets were registered with precision of 0.2 mu s steps. Experimental measurements were made using five 29 mm Edwards-Duromedics(TM) including three original specification (EDOS) and two modified specification (EDMS), and two 29 mm St Jude Medical(R) MHVs. The testing valve was installed in the mitral position of a physiologic pulsatile mock circulatory flow loop using water-glycerine solution as the testing fluid. Each valve was tested by: (1) direct mounting the valve on metal washers, and (2) mounting the valve with its sewing ring. Experiments were carried out al pulse rates of 70, 90, and 120 beats min(-1), with the corresponding cardiac output of 5, 6, and 7.5 litres min(-1), and maximum left ventricular pressure gradients (dp/dt) of 1,800, 3,000 and 5,600 mm Hg s(-1), respectively. The maxi,mum leaflet closing velocity of each of the tested valve types are presented. The difference in leaflet closing movements between the direct rigid mounting and the sewing ring mounting are discussed. The details of the laser sweeping technique are presented.
引用
收藏
页码:458 / 464
页数:7
相关论文
共 11 条
[1]  
ABDALLAH SA, 1983, T AM SOC ART INT ORG, V29, P296
[2]  
ADRAIN RJ, 1983, FLUID MECHANICS MEAS, P155
[3]  
BOKROS JC, 1972, CHEM PHYS CARBON, P103
[4]   STATIC AND DYNAMIC STRESSES DURING VALVE CLOSURE OF A BILEAFLET MECHANICAL HEART-VALVE PROSTHESIS [J].
CHIANG, TH ;
LAM, H ;
QUIJANO, R ;
DONHAM, R ;
GILLIAM, P ;
HEINZ, LA .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (12) :781-788
[5]   LASER ASSESSMENT OF LEAFLET CLOSING MOTION IN PROSTHETIC HEART-VALVES [J].
GUO, GX ;
XU, CC ;
HWANG, NHC .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :477-481
[6]  
GUO GX, 1990, T AM SOC ART INT ORG, V36, P529
[7]  
KAFESJIAN R, 1989, JUN P INT S HEART VA, P509
[8]  
KLEPETKO W, 1989, J THORAC CARDIOV SUR, V97, P90
[9]  
KOYMEN H, 1986, 8TH P IEEE ANN C ENG, P136
[10]  
LEUER L, 1986, JAN P INT S ART ORG, P5