DRIFT DIAGNOSTICS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY

被引:42
作者
CARRE, M [1 ]
POUSSEL, E [1 ]
MERMET, JM [1 ]
机构
[1] UNIV LYON 1,SCI ANALYT LAB,F-69622 VILLEURBANNE,FRANCE
关键词
INDUCTIVELY COUPLED PLASMA; ATOMIC EMISSION SPECTROMETRY; DRIFT;
D O I
10.1039/ja9920700791
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple experiment is described to carry out drift diagnostics in inductively coupled plasma atomic emission spectrometry. It is based on the behaviour of the Ba II 455.403, Zn II 206.200 and Ar I 404.442 nm line intensities as a function of the power, the solution uptake rate and the carrier gas flow rate. Additional information on the atomization and ionization conditions is obtained from the Mg II 280.270 nm to Mg I 285.213 nm line intensity ratio. Methods to verify the presence of a drift over a limited period of time are discussed. Several examples illustrate the feasibility of these diagnostics. A conclusion is drawn about the selection of internal standards for drift correction.
引用
收藏
页码:791 / 797
页数:7
相关论文
共 16 条
[1]   ELIMINATION OF INTERFERENCES AND EFFECTS FROM DRIFT IN WORKING PARAMETERS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY BY USING A COMBINATION OF THE GENERALIZED STANDARD ADDITION METHOD AND THE GENERALIZED INTERNAL REFERENCE METHOD [J].
ALAMMAR, AS ;
HAMID, HA ;
RASHID, BH .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (4-5) :359-366
[2]   MEASURING DETECTION LIMITS IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY .2. EXPERIMENTAL-DATA AND THEIR INTERPRETATION [J].
BOUMANS, PWJM ;
IVALDI, JC ;
SLAVIN, W .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (05) :641-665
[3]   CROSS-FLOW NEBULIZERS AND TESTING PROCEDURES FOR INDUCTIVELY COUPLED PLASMA NEBULIZERS [J].
IVALDI, JC ;
SLAVIN, W .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (05) :359-364
[4]   CORRECTION FOR DRIFT BY INTERNAL REFERENCE METHODS IN INDUCTIVELY COUPLED PLASMA SIMULTANEOUS MULTIELEMENT ANALYSIS [J].
LORBER, A ;
GOLDBART, Z ;
ELDAN, M .
ANALYTICAL CHEMISTRY, 1984, 56 (01) :43-48
[5]   APPLICATION OF THE GENERALIZED INTERNAL REFERENCE METHOD FOR THE CHARACTERIZATION OF PARAMETERS CAUSING DRIFT IN INDUCTIVELY-COUPLED PLASMA EMISSION-SPECTROMETRY [J].
LORBER, A ;
GOLDBART, Z .
ANALYTICA CHIMICA ACTA, 1984, 161 (JUL) :163-173
[6]  
MARK H, 1991, STATISTICS SPECTROSC, pCH7
[7]   AN EMISSION-BASED FEEDBACK-SYSTEM FOR STABILIZING AN INDUCTIVELY COUPLED PLASMA [J].
MARKS, MA ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1988, 42 (02) :277-280
[8]  
MASSART DL, 1988, CHEMOMETRICS TXB, pCH6
[9]   USE OF MAGNESIUM AS A TEST ELEMENT FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY DIAGNOSTICS [J].
MERMET, JM .
ANALYTICA CHIMICA ACTA, 1991, 250 (01) :85-94