Diagnostic value of two modes of cone-beam computed tomography in evaluation of simulated external root resorption: an in vitro study

被引:29
作者
Dalili, Zahra [1 ]
Taramsari, Mehran [2 ]
Mehr, Seyed Zoheir Mousavi [1 ]
Salamat, Fatemeh [3 ]
机构
[1] Guilan Univ Med Sci, Dept Maxillofacial Radiol, Dent Sch, Rasht, Iran
[2] Guilan Univ Med Sci, Dept Endodont, Dent Sch, Rasht, Iran
[3] Guilan Univ Med Sci, Dept Epidemiol, Rasht, Iran
关键词
Cone Beam Computed Tomography; Root Resorption; Diagnosis;
D O I
10.5624/isd.2012.42.1.19
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose : Field of view and voxel resolution of cone beam computed tomography (CBCT) might affect the diagnostic capability. This study was performed to compare between the standard and HiRes zoom modes in the diagnosis of external root resorption (ERR) using CBCT. Materials and Methods : Sixty three small cavities (0.25 mm depth and 0.5 mm diameter) were simulated on the buccal, lingual, and proximal surfaces at three different levels of 16 roots of teeth. After covering the root with nail varnish, the roots were inserted in the sockets and the model was placed in a water-containing lacuna. CBCT scans were taken in both standard and HiRes zoom modes using NewTom VG (QR srl Company, Verona, Italy). Then, an observer assessed the images to determine the presence or absence of the cavities. This process was repeated by increasing the size and depth of cavities to 0.5 mm depth and 1 mm diameter. Data were analyzed by McNemar test. The sensitivity, specificity, positive predictive value, negative predictive value, and likelihood ratio in evaluation of the simulated cavities were calculated. Results : There was a significant difference between the two imaging modes in diagnosing the shallow cavities (p=0.02). The sensitivity of the standard zoom in detecting the shallow cavities was lower than that of the HiRes zoom. The likelihood ratio of the HiRes zoom was higher in the diagnosis of both cavity types. Conclusion : This study suggested that a smaller voxel size in the HiRes zoom mode of CBCT is preferred for diagnosis of ERR.
引用
收藏
页码:19 / 24
页数:6
相关论文
共 16 条
[1]  
ANDREASEN JO, 1987, ENDOD DENT TRAUMATOL, V3, P269
[2]   Film and digital radiography for detection of simulated root resorption cavities [J].
Borg, E ;
Kallqvist, A ;
Grondahl, K ;
Grondahl, HG .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1998, 86 (01) :110-114
[3]   EXTERNAL ROOT RESORPTION - AN EXPERIMENTAL RADIOGRAPHIC EVALUATION [J].
CHAPNICK, L .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1989, 67 (05) :578-582
[4]   Diagnostic ability of computed tomography to evaluate external root resorption in vitro [J].
da Silveira, H. L. D. ;
Silveira, H. E. D. ;
Liedke, G. S. ;
Lermen, C. A. ;
dos Santos, R. B. ;
de Figuleiredo, J. A. P. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (07) :393-396
[5]   Method to Evaluate Inflammatory Root Resorption by Using Cone Beam Computed Tomography [J].
Estrela, Carlos ;
Bueno, Mike Reis ;
Goncalves De Alencar, Ana Helena ;
Mattar, Rinaldo ;
Valladares Neto, Jose ;
Azevedo, Bruno Correa ;
De Araujo Estrela, Cyntia Rodrigues .
JOURNAL OF ENDODONTICS, 2009, 35 (11) :1491-1497
[6]  
Goldberg F, 1998, ENDOD DENT TRAUMATOL, V14, P133
[7]   Detection and size differentiation of simulated tooth root defects using flat-panel volume computerized tomography (fpVCT) [J].
Hahn, Wolfram ;
Fricke-Zech, Susanne ;
Fricke, Julia ;
Gruber, Rudolf M. ;
Dullin, Christian ;
Zapf, Antonia ;
Hannig, Christian ;
Kubein-Meesenburg, Dietmar ;
Sadat-Khonsari, Reza .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 107 (02) :272-278
[8]  
HINTZE H, 1992, ENDOD DENT TRAUMATOL, V8, P149
[9]  
Hulsmann M, 2009, PROBLEMS ENDODONTICS, P421
[10]   Computed tomography as a diagnostic aid for extracanal invasive resorption [J].
Kim, E ;
Kim, KD ;
Roh, BD ;
Cho, YS ;
Lee, SJ .
JOURNAL OF ENDODONTICS, 2003, 29 (07) :463-465