SIMPLE, RELIABLE, AND SENSITIVE INTERFEROMETER FOR THE MEASUREMENT OF THE REFRACTIVE-INDEX OF LIQUIDS AS A FUNCTION OF TEMPERATURE

被引:17
作者
BEYSENS, D
机构
[1] Service de Physique du Solide et de Résonance Magnétique, Commissariat a l'Energie Atomique, 91190 Gif-fur-Yvette
关键词
D O I
10.1063/1.1135861
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe a very simple and inexpensive interferometer setup which provides high accuracy measurements of the refractive index (n) of liquids with temperature. A standard deviation σ (n) =5.4×10-7 has been observed in isooctane liquid when fitting n to a linear temperature variation.
引用
收藏
页码:509 / 510
页数:2
相关论文
共 2 条
  • [1] GRAY DE, 1957, AM I PHYSICS HDB
  • [2] TOURNARIE M, 1969, J PHYS-PARIS, V30, P47