FOURIER CODED-APERTURE IMAGING IN NUCLEAR-MEDICINE

被引:0
|
作者
CHOU, C
机构
[1] Natl Yang Ming Medical Coll, Taipei
关键词
CODED APERTURE; GAMMA-RAY IMAGING; MOIRE; TOMOGRAPHY;
D O I
10.1049/ip-smt:19949767
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A Fourier aperture (FA) is composed of two sinusoidal gratings in contact. When the two gratings rotate on the same axis in equal angles but opposite directions a clear pattern of difference moire fringes is formed, the moire frequency depending on the angle between the two gratings. This aperture of moire fringes is used as a variable frequency grating which is able to scan the object spectrum in the frequency domain. The summation-filtered back projection algorithm with three different apodising functions was introduced in the reconstruction. A Tc-99m gamma ray source and a 1 in-diameter NaI scintillation detector were used in the experiment. The thyroid phantom images showed that spatial resolution of FA agrees with the theoretical prediction. Three dimensional image capability is also discussed. The matrix inversion algorithm was used in the computer simulation of 3-D images. Although the depth resolution of a FA is limited by the restricted view angle of the detector, two plane objects free from noise at different depths were reconstructed.
引用
收藏
页码:179 / 184
页数:6
相关论文
共 50 条
  • [21] QUANTUM-NOISE OF FOURIER-CODED APERTURE IMAGING-SYSTEM
    CHOU, C
    KING, HJ
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1994, 33 (4A): : 2072 - 2078
  • [22] Spectral Image Classification From Optimal Coded-Aperture Compressive Measurements
    Ramirez, Ana
    Arguello, Henry
    Arce, Gonzalo R.
    Sadler, Brian M.
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2014, 52 (06): : 3299 - 3309
  • [23] Compressive Coded Aperture Imaging
    Marcia, Roummel F.
    Harmany, Zachary T.
    Willett, Rebecca M.
    COMPUTATIONAL IMAGING VII, 2009, 7246
  • [24] New principles in Nuclear Medicine Imaging: A Full Aperture Stereoscopic Imaging Technique
    Strocovsky, Sergio G.
    Otero, Dino
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 3630 - 3633
  • [25] Method of sparse-view coded-aperture x-ray diffraction tomography
    Liang, Kaichao
    Zhang, Li
    Xing, Yuxiang
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (06)
  • [26] Complementary Sequences for Coded Aperture Imaging
    Cristea, Paul Dan
    Tuduce, Rodica
    Cornelis, Jan
    PROCEEDINGS ELMAR-2008, VOLS 1 AND 2, 2008, : 53 - +
  • [27] A new contrast-to-noise ratio for image quality characterization of a coded-aperture γ camera
    Zhu, Balin
    Wang, Zhonghai
    Gao, Tai
    Chen, Qi
    Huang, Qichang
    Mao, Xiaoqin
    Zhao, Chaoyang
    Tao, Mingrui
    Qi, Peiwen
    Zhao, Qianru
    Li, Yulong
    Zhou, Rong
    Yang, Chaowen
    APPLIED RADIATION AND ISOTOPES, 2021, 170
  • [28] Real time depth-of-interaction correction for coded-aperture, gamma-ray images
    Ziock, K. P.
    Blackston, M. A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 916 : 56 - 65
  • [29] Hybrid coded aperture and Compton imaging using an active mask
    Schultz, L. J.
    Wallace, M. S.
    Galassi, M. C.
    Hoover, A. S.
    Mocko, M.
    Palmer, D. M.
    Tornga, S. R.
    Kippen, R. M.
    Hynes, M. V.
    Toolin, M. J.
    Harris, B.
    McElroy, J. E.
    Wakeford, D.
    Lanza, R. C.
    Horn, B. K. P.
    Wehe, D. K.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (02) : 267 - 274
  • [30] Row-Column Readout Method to Mitigate Radiographic-Image Blurring From Multipixel Events in a Coded-Aperture Imaging System
    Boo, Jihwan
    Hammig, Mark D.
    Jeong, Manhee
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2021, 68 (05) : 1175 - 1183