PREDICTION OF RETENTION TIMES OF THE NUCLEOSIDES AND BASES ON REVERSE PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY DURING GRADIENT ELUTION

被引:44
作者
HARTWICK, RA [1 ]
GRILL, CM [1 ]
BROWN, PR [1 ]
机构
[1] UNIV RHODE ISL,DEPT CHEM,KINGSTON,RI 02881
关键词
D O I
10.1021/ac50037a016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Through the use of relationships between in k' and composition of the mobile phase, it is possible to predict to within about 5% the retention times of the nucleosides and bases under gradient elution conditions. Through the use of equations, optimal separation of the compounds hypoxanthine, uridine, xanthine, and six other compounds was achieved using reverse phase (C18) columns. The equations predicted the gradient conditions needed to separate these compounds in minimum time with maximum resolution. © 1979, American Chemical Society. All rights reserved.
引用
收藏
页码:34 / 38
页数:5
相关论文
共 10 条
[1]   CHROMATOGRAPHIC REPRODUCIBILITY IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC GRADIENT ELUTION [J].
ABBOTT, SR ;
BERG, JR ;
ACHENER, P ;
STEVENSON, RL .
JOURNAL OF CHROMATOGRAPHY, 1976, 126 (NOV3) :421-437
[2]   COMPARISON OF SOME PACKINGS FOR REVERSED-PHASE HIGH-PERFORMANCE LIQUID-SOLID CHROMATOGRAPHY .1. SOME ANALYTICAL CONSIDERATIONS [J].
COLIN, H ;
WARD, N ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHY, 1978, 149 (FEB) :169-197
[3]  
DAVIS GE, 1977, CLIN CHEM, V23, P1427
[4]   EVALUATION OF MICROPARTICLE CHEMICALLY BONDED REVERSED-PHASE PACKINGS IN HIGH-PRESSURE LIQUID-CHROMATOGRAPHIC ANALYSIS OF NUCLEOSIDES AND THEIR BASES [J].
HARTWICK, RA ;
BROWN, PR .
JOURNAL OF CHROMATOGRAPHY, 1976, 126 (NOV3) :679-691
[5]   GRADIENT ELUTION IN LIQUID-CHROMATOGRAPHY .4. VERIFICATION OF THEORETICAL RELATIONSHIPS FOR ELUTION CHARACTERISTICS (RETENTION VOLUME, BAND WIDTH, RESOLUTION, PLATE NUMBER) IN ADSORPTION CHROMATOGRAPHY ON SILICA USING STEPWISE AND COMBINED (STEPWISE-GRADIENT) ELUTION AND SOME CONSIDERATIONS CONCERNING SOLVENT-PROGRAMMED AND ISOCRATIC ELUTION CHROMATOGRAPHYC [J].
JANDERA, P ;
CHURACEK, J .
JOURNAL OF CHROMATOGRAPHY, 1975, 104 (01) :9-21
[6]   GRADIENT ELUTION IN LIQUID-CHROMATOGRAPHY .3. VERIFICATION OF THEORETICAL RELATIONSHIPS FOR ELUTION CHARACTERISTICS (RETENTION VOLUME, BAND WIDTH AND RESOLUTION) IN ISOCRATIC AND GRADIENT ELUTION CHROMATOGRAPHY ON SILICA [J].
JANDERA, P ;
CHURACEK, J .
JOURNAL OF CHROMATOGRAPHY, 1974, 93 (01) :17-39
[7]   GRADIENT ELUTION IN LIQUID-CHROMATOGRAPHY .1. INFLUENCE OF COMPOSITION OF MOBILE PHASE ON CAPACITY RATIO (RETENTION VOLUME, BAND WIDTH, AND RESOLUTION) IN ISOCRATIC ELUTION - THEORETICAL CONSIDERATIONS [J].
JANDERA, P ;
CHURACEK, J .
JOURNAL OF CHROMATOGRAPHY, 1974, 91 (APR24) :207-221
[8]   SELECTIVITY IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY USING CHEMICALLY BONDED STATIONARY PHASES [J].
LOCKE, DC .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1974, 12 (08) :433-437
[9]   GRADIENT SELECTION IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
SCHOENMAKERS, PJ ;
BILLIET, HAH ;
TIJSSEN, R ;
DEGALAN, L .
JOURNAL OF CHROMATOGRAPHY, 1978, 149 (FEB) :519-537
[10]   SOLUTE-SOLVENT INTERACTIONS ON SURFACE OF SILICA-GEL [J].
SCOTT, RPW ;
KUCERA, P .
JOURNAL OF CHROMATOGRAPHY, 1978, 149 (FEB) :93-110