QUALITY-CONTROL OF INHOMOGENEITY CORRECTION ALGORITHMS USED IN TREATMENT PLANNING SYSTEMS

被引:16
|
作者
KAPPAS, C [1 ]
ROSENWALD, JC [1 ]
机构
[1] INST PASTEUR,SERV PHYS MED,PARIS,FRANCE
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1995年 / 32卷 / 03期
关键词
RADIOTHERAPY; QUALITY CONTROL; INHOMOGENEITIES; RADIOTHERAPY TREATMENT PLANNING;
D O I
10.1016/0360-3016(94)00474-Y
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This quality control program has been carried out under the auspices of S.F.P.H. (Societe Francaise des Physiciens d'Hopital), to evaluate the performances of radiotherapy treatment planning systems (RTPS) used by different institutions, The aim of this Quality Assurance Programme was: (a) to set up a methodology to assess globally the capability of a given system to perform inhomogeneity corrections in the irradiated medium with external photon beams; (b) to analyze the limitations of the algorithms presently used and especially the two-dimensional (2D) dose calculation possibilities; (c) to check, on a number of systems in clinical use, the validity of the method and the variation of the results as compared to measurements used as reference. Methods and Materials: Phantom (lung equivalent material placed into polystyrene) measurements, using cobalt-60 radiation, were carried out by the authors. The phantoms were circulated among the participating institutes to be scanned, and used as input to the treatment planning computer. Results: Ten systems were tested in this study, using seven different inhomogeneity correction algorithms implemented in nine different TPS; four of these algorithms are used in a pixel by pixel basis and five of them in a contour basis. Significant discrepancies or inconsistencies have been observed even for sophisticated models supposed to be mostly accurate. Conclusion: The proposed tests and the experimental data provided are very useful as part of a quality-control program. They should be included in the initial extensive validation of TPS before starting clinical use, and should be repeated at regular intervals and at each updating of the program. They have the merit of including the whole procedure, from patient data acquisition to dose distribution printout.
引用
收藏
页码:847 / 858
页数:12
相关论文
共 37 条
  • [21] New automatic quality control methods for geometrical treatment planning system tools in external conformal radiotherapy
    Denis, Eloise
    Beaumont, Stephane
    Guedon, JeanPierre
    Torfeh, Tarraf
    Normand, Nicolas
    Ailleres, Norbert
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [22] Dosimetric comparison of treatment planning systems using collapsed cone convolution and pencil beam algorithms
    Elcim, Yelda
    Dirican, Bahar
    Yavas, Omer
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2016, 15 (04) : 364 - 377
  • [23] RADIATION-THERAPY QUALITY-CONTROL IN A CLINICAL-TRIAL OF ADJUVANT POSTOPERATIVE TREATMENT FOR RECTAL-CANCER
    MARTENSON, JA
    URIAS, R
    SMALLEY, SR
    COIA, LR
    TEPPER, JE
    ROTMAN, M
    RICH, TA
    OCONNELL, MJ
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (01): : 51 - 55
  • [24] QUALITY OF INTRAORAL RADIOGRAPHS USED FOR PROSTHODONTIC TREATMENT PLANNING BY GENERAL DENTISTS IN THE PUBLIC DENTAL-HEALTH SERVICE
    SVENSON, B
    ERIKSSON, T
    KRONSTROM, M
    PALMQVIST, S
    SWEDISH DENTAL JOURNAL, 1995, 19 (1-2) : 47 - 54
  • [25] COMP Report: CPQR technical quality control guidelines for use of positron emission tomography/computed tomography in radiation treatment planning
    Klein, Ran
    Oliver, Mike
    La Russa, Dan
    Agapito, John
    Gaede, Stewart
    Bissonnette, JeanPierre
    Rahmim, Arman
    Uribe, Carlos
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [26] Integrated production planning and quality control for linear production systems under uncertainties of cycle time and finished product quality
    Ben-Ammar, Oussama
    Bettayeb, Belgacem
    Dolgui, Alexandre
    INTERNATIONAL JOURNAL OF PRODUCTION RESEARCH, 2020, 58 (04) : 1144 - 1160
  • [27] Statistical control process to compare and rank treatment plans in radiation oncology: impact of heterogeneity correction on treatment planning in lung cancer
    Chaikh, Abdulhamid
    Balosso, Jacques
    TRANSLATIONAL LUNG CANCER RESEARCH, 2016, 5 (06) : 688 - 694
  • [28] Patient-related quality assurance with different combinations of treatment planning systems, techniques, and machines
    Steiniger, Beatrice
    Berger, Rene
    Eilzer, Sabine
    Kornhuber, Christine
    Lorenz, Kathleen
    Peil, Torsten
    Reiffenstuhl, Carsten
    Schilz, Johannes
    Schroeder, Dirk
    Schwedas, Michael
    Pensold, Stephanie
    Walke, Mathias
    Weibert, Kirsten
    Wolf, Ulrich
    Wiezorek, Tilo
    STRAHLENTHERAPIE UND ONKOLOGIE, 2017, 193 (01) : 46 - 54
  • [29] QUALITY-CONTROL OF VEGETABLE OIL METHYL-ESTERS USED AS DIESEL FUEL SUBSTITUTES - QUANTITATIVE-DETERMINATION OF MONOGLYCERIDES, DIGLYCERIDES, AND TRIGLYCERIDES BY CAPILLARY GC
    PLANK, C
    LORBEER, E
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1992, 15 (09): : 609 - 612
  • [30] Quality-control tests for OC4, OC5 and NIR-red satellite chlorophyll-a algorithms applied to coastal waters
    Lavigne, H.
    Van der Zande, D.
    Ruddick, K.
    Dos Santos, J. F. Cardoso
    Gohin, F.
    Brotas, V.
    Kratzer, S.
    REMOTE SENSING OF ENVIRONMENT, 2021, 255