DETERMINATION AND APPLICATION OF ION-SELECTIVE ELECTRODE MODEL PARAMETERS USING FLOW-INJECTION AND SIMPLEX OPTIMIZATION

被引:153
作者
DEVITERI, FJS [1 ]
DIAMOND, D [1 ]
机构
[1] DUBLIN CITY UNIV,SCH CHEM SCI,DUBLIN 9,IRELAND
关键词
SENSOR ARRAYS; ION-SELECTIVE ELECTRODES; AMMONIUM; SODIUM; POTASSIUM; CALCIUM; SELECTIVITY COEFFICIENTS; SIMPLEX OPTIMIZATION; MULTIVARIATE CALIBRATION;
D O I
10.1039/an9941900749
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A modified form of the Nikolskii-Eisenman equation is proposed in which the power dependence of the interfering ion activity is removed. Selectivity is determined for poly(vinyl chloride) membrane electrodes (Na+, K+, Ca2+ and NH4+) in terms of constants (which are valid only for the particular experimental design and conditions used) rather than conventional coefficients. These constants and other electrode parameters, such as cell constant and slope, are measured by means of an ion-selective array approach coupled with flow injection. Each electrode in the array is characterized by means of a fractional factorial calibration design and simplex optimization of the electrode parameters. The applicability of the resulting values for these parameters is demonstrated through the determination of unknowns by direct solution of the system of modified Nikolskii-Eisenman equations describing the array response. The approach is suitable for use in real-time monitoring applications where batch calibration techniques cannot easily be implemented.
引用
收藏
页码:749 / 758
页数:10
相关论文
共 23 条
  • [1] ADAMS MI, 1992, PRACTICAL GUIDE CHEM, P205
  • [2] SPARINGLY SELECTIVE ION-SELECTIVE ELECTRODE ARRAYS FOR MULTICOMPONENT ANALYSIS
    BEEBE, K
    UERZ, D
    SANDIFER, J
    KOWALSKI, B
    [J]. ANALYTICAL CHEMISTRY, 1988, 60 (01) : 66 - 71
  • [3] NONLINEAR CALIBRATION USING PROJECTION PURSUIT REGRESSION - APPLICATION TO AN ARRAY OF ION-SELECTIVE ELECTRODES
    BEEBE, KR
    KOWALSKI, BR
    [J]. ANALYTICAL CHEMISTRY, 1988, 60 (20) : 2273 - 2278
  • [4] SODIUM-SELECTIVE POLYMERIC MEMBRANE ELECTRODES BASED ON CALIX[4]ARENE IONOPHORES
    CADOGAN, AM
    DIAMOND, D
    SMYTH, MR
    DEASY, M
    MCKERVEY, MA
    HARRIS, SJ
    [J]. ANALYST, 1989, 114 (12) : 1551 - 1554
  • [5] COVINGTON AK, 1979, ION SELECTIVE ELECTR, P18
  • [6] OPTIMIZATION OF POLYVINYL-CHLORIDE) MATRIX MEMBRANE ION-SELECTIVE ELECTRODES FOR AMMONIUM-IONS
    DAVIES, OG
    MOODY, GJ
    THOMAS, JDR
    [J]. ANALYST, 1988, 113 (03) : 497 - 500
  • [7] AMMONIUM DETECTION USING AN ION-SELECTIVE ELECTRODE ARRAY IN FLOW-INJECTION ANALYSIS
    DEVITERI, FJS
    DIAMOND, D
    [J]. ELECTROANALYSIS, 1994, 6 (01) : 9 - 16
  • [8] MULTICOMPONENT BATCH-INJECTION ANALYSIS USING AN ARRAY OF ION-SELECTIVE ELECTRODES
    DIAMOND, D
    LU, JM
    CHEN, Q
    WANG, J
    [J]. ANALYTICA CHIMICA ACTA, 1993, 281 (03) : 629 - 635
  • [9] A SODIUM ION-SELECTIVE ELECTRODE BASED ON METHYL PARA-TERT-BUTYLCALIX[4]ARYL ACETATE AS THE IONOPHORE
    DIAMOND, D
    SVEHLA, G
    SEWARD, EM
    MCKERVEY, MA
    [J]. ANALYTICA CHIMICA ACTA, 1988, 204 (1-2) : 223 - 231
  • [10] ROBUST ESTIMATION OF SELECTIVITY COEFFICIENTS USING MULTIVARIATE CALIBRATION OF ION-SELECTIVE ELECTRODE ARRAYS
    DIAMOND, D
    FORSTER, RJ
    [J]. ANALYTICA CHIMICA ACTA, 1993, 276 (01) : 75 - 86