BONE-FRACTURE HEALING USING A CAPACITATIVELY COUPLED RF FIELD

被引:9
作者
BEHARI, J [1 ]
ARYA, VP [1 ]
ALEX, ZC [1 ]
机构
[1] JAWAHARLAL NEHRU UNIV,CTR SCI & INSTRUMENTAT,NEW DELHI 110067,INDIA
来源
JOURNAL OF BIOELECTRICITY | 1991年 / 10卷 / 1-2期
关键词
D O I
10.3109/15368379109031409
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A new pulsed radio frequency electric field pattern evolved from bone studies was adopted to accelerate fracture healing in rats. The output of a specially designed signal generator was capacitatively coupled to the fracture site using a pair of stainless steel electrodes. In a series of experiments performed on rats, fractures were induced in the femoral shafts and electrical stimulations were applied to one leg. Bone mass formed in the gap was estimated by measurement of the cortical thickness and by ultrasonic attenuation. We found that the stimulated side showed greater bone mass than the contralateral control. We suggest that this device offers a simple and reliable method of accelerated healing, immediately after fracture.
引用
收藏
页码:231 / 239
页数:9
相关论文
共 17 条
[1]  
BARANOWSKI T J JR, 1983, Journal of Orthopaedic Research, V1, P120, DOI 10.1002/jor.1100010202
[2]  
BASSETT CAL, 1989, CRIT REV BIOMED ENG, V17, P451
[3]  
BASSETT CAL, 1978, METABOLIC SURGERY, P255
[4]  
BEHARI J, 1990, CRITICAL REV BIOENGI, P235
[5]  
BEHARI J, 1986, Patent No. 438
[6]  
BLACK J, 1979, ELECTRICAL PROPERTIE, P215
[7]   TREATMENT OF RECALCITRANT NON-UNION WITH A CAPACITIVELY COUPLED ELECTRICAL-FIELD - A PRELIMINARY-REPORT [J].
BRIGHTON, CT ;
POLLACK, SR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (04) :577-585
[8]  
BRIGHTON CT, 1984, ORTHOP CLIN N AM, V15, P33
[9]  
Brighton CT., 1975, CLIN ORTHOP RELAT R, DOI [10.1097/00003086-197503000-00033, DOI 10.1097/00003086-197503000-00033]
[10]  
CONNOLLY JF, 1981, CLIN ORTHOP RELAT R, V161, P39