EVALUATION OF THE L-CLONE LEGIONELLA-PNEUMOPHILA SEROGROUP-1 URINE ANTIGEN LATEX TEST

被引:16
|
作者
LELAND, DS [1 ]
KOHLER, RB [1 ]
机构
[1] INDIANA UNIV,MED CTR,DEPT MED,INDIANAPOLIS,IN 46202
关键词
D O I
10.1128/JCM.29.10.2220-2223.1991
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The purpose of this study was to evaluate the L-CLONE Legionella pneumophila Serogroup 1 Urine Antigen Latex Test (Access Medical Systems, Inc., Branford, Conn.) for detection of Legionella antigen in urine. A total of 481 frozen urine samples previously tested by an in-house solid-phase radioimmunoassay (RIA) was thawed and retested by using L-CLONE. Included in this sample were 140 RIA-positive samples from culture-positive or serologically confirmed cases of legionellosis and 341 RIA-negative samples from patients with non-Legionella respiratory disease or bacteriuria. The original RIA test result was accepted as the true value. L-CLONE correctly identified 76 of 140 (54%) known positive samples. False-negative results could not be attributed to a low Legionella antigen concentration or to a Legionella antigen subgroup. L-CLONE correctly identified 252 of 341 (74%) known negative samples. False-positive results were experienced in all groups of negative samples, regardless of the patients' underlying diseases. A total of 141 fresh urine samples was tested; all were Legionella antigen negative by RIA. L-CLONE provided 86% specificity. The sensitivity of the L-CLONE in testing fresh urine samples could not be evaluated because of the lack of Legionella antigen RIA-positive samples.