CANINE AND FELINE HEMATOLOGY ANALYSIS - COMPARATIVE PERFORMANCE OF TECHNICON H-ASTERISK-1 AND AVL MS8 VET ANALYZERS

被引:5
作者
BYRNE, RF
ANDREWS, CM
LIBRETTO, SE
MIFSUD, CV
机构
[1] GLAXO GRP RES LTD,RES & DEV,DIV MED SAFETY EVALUAT,WARE,HERTS,ENGLAND
[2] W BAR VET HOSP,BANBURY,OXON,ENGLAND
来源
COMPARATIVE HAEMATOLOGY INTERNATIONAL | 1994年 / 4卷 / 04期
关键词
CAT; DOG; IMPEDENCE ANALYSIS; FLOW CYTOMETRIC ANALYSIS; PSEUDOLEUCOCYTOSIS; PSEUDOTHROMBOCYTOPENIA; MACROCYTOSIS; POSTAL TRANSPORTATION;
D O I
10.1007/BF00185176
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The performance of a prototype AVL MS8 VET impedance haematology analyser was compared with that of a Technicon H*1 flow cytometry haematology analyser using blood from dogs and cats. Analysis was performed with the AVL MS8 VET on the day of blood collection and with the Technicon H*1 on the following day. Differences were noted in feline leucocyte and platelet counts and in canine and feline mean cell volume and mean cell haemoglobin concentration between analyses. The results indicate that the AVL MS8 VET is a reliable analyser for blood samples from dogs but may not be for those from cats. Attention is drawn to the importance of considering the type of analyser, calibration of the analyser, time of analysis after blood collection (effect of postal delay) and the effect of anticoagulants.
引用
收藏
页码:212 / 217
页数:6
相关论文
共 18 条
[1]  
BRITTIN GM, 1969, AM J CLIN PATHOL, V52, P679
[2]  
Davies D. T., 1991, Comparative Haematology International, V1, P91, DOI 10.1007/BF00422879
[3]   COMPARISON OF THE HEMATOLOGIC EFFECTS OF 2-BUTOXYETHANOL USING 2 TYPES OF HEMATOLOGY ANALYZERS [J].
GHANAYEM, BI ;
WARD, SM ;
BLAIR, PC ;
MATTHEWS, HB .
TOXICOLOGY AND APPLIED PHARMACOLOGY, 1990, 106 (02) :341-345
[4]  
HINCHLIFFE RF, 1992, CLIN LAB HAEMATOL, V14, P268
[5]  
Jain N. C., 1986, SCHALMS VET HAEMATOL
[6]  
KIMBALL JP, 1992, 5 P C INT SOC AN BIO
[7]  
LEVIN J, 1983, J LAB CLIN MED, V101, P295
[8]   UNPREDICTABLE EFFECTS OF K3 EDTA ON MEAN PLATELET VOLUME [J].
LIPPI, U ;
SCHINELLA, M ;
MODENA, N ;
NICOLI, M .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1987, 87 (03) :391-393
[9]   ADVANTAGES OF A NEW ANTICOAGULANT IN ROUTINE HEMATOLOGY ON THE COULTER-COUNTER S-PLUS STKR ANALYZER [J].
LIPPI, U ;
SCHINELLA, M ;
MODENA, N ;
NICOLI, M ;
LIPPI, G .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1990, 93 (06) :760-764
[10]  
ONDER O, 1980, BLOOD, V56, P177