INTENSIFIED CHARGE COUPLED DEVICE-BASED FIBEROPTIC MONITOR FOR RAPID REMOTE SURFACE-ENHANCED RAMAN-SCATTERING SENSING

被引:37
作者
ALARIE, JP [1 ]
STOKES, DL [1 ]
SUTHERLAND, WS [1 ]
EDWARDS, AC [1 ]
VO-DINH, T [1 ]
机构
[1] OAK RIDGE NATL LAB, ADV MONITORING DEV GRP, OAK RIDGE, TN 37831 USA
关键词
INTENSIFIED CHARGE COUPLED DEVICE; SURFACE-ENHANCED RAMAN; FIBEROPTIC PROBE; REMOTE SENSING;
D O I
10.1366/0003702924926736
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper describes the development of an intensified charge coupled device (ICCD)-based fiber-optic monitor for remote Raman and surface-enhanced Raman (SERS) sensing. Both Raman and SERS data were obtained with the use of a fiber-optic probe design incorporating 20-m optical fibers carrying the Raman signal. Spectra were obtained in 5 milliseconds for Raman and 9 ms for SERS. The proposed system could be used for a highly sensitive portable Raman system for rapid and remote chemical sensing.
引用
收藏
页码:1608 / 1612
页数:5
相关论文
共 28 条
[1]   SILVER-COATED FUMED SILICA AS A SUBSTRATE MATERIAL FOR SURFACE-ENHANCED RAMAN-SCATTERING [J].
ALAK, AM ;
VO-DINH, T .
ANALYTICAL CHEMISTRY, 1989, 61 (07) :656-660
[2]   NEAR-INFRARED RAMAN-SPECTROSCOPY OF LIQUIDS AND SOLIDS WITH A FIBEROPTIC SAMPLER, DIODE-LASER, AND CCD DETECTOR [J].
ALLRED, CD ;
MCCREERY, RL .
APPLIED SPECTROSCOPY, 1990, 44 (07) :1229-1231
[3]   FIBEROPTIC REMOTE SENSOR FOR INSITU SURFACE-ENHANCED RAMAN-SCATTERING ANALYSIS [J].
BELLO, JM ;
NARAYANAN, VA ;
STOKES, DL ;
VO-DINH, T .
ANALYTICAL CHEMISTRY, 1990, 62 (22) :2437-2441
[4]   SURFACE-ENHANCED RAMAN-SCATTERING FIBEROPTIC SENSOR [J].
BELLO, JM ;
VO-DINH, T .
APPLIED SPECTROSCOPY, 1990, 44 (01) :63-69
[5]   TITANIUM-DIOXIDE BASED SUBSTRATE FOR OPTICAL MONITORS IN SURFACE-ENHANCED RAMAN-SCATTERING ANALYSIS [J].
BELLO, JM ;
STOKES, DL ;
VO-DINH, T .
ANALYTICAL CHEMISTRY, 1989, 61 (15) :1779-1783
[6]   SILVER-COATED ALUMINA AS A NEW MEDIUM FOR SURFACED-ENHANCED RAMAN-SCATTERING ANALYSIS [J].
BELLO, JM ;
STOKES, DL ;
VO-DINH, T .
APPLIED SPECTROSCOPY, 1989, 43 (08) :1325-1330
[7]   SPECTROCHEMICAL MEASUREMENTS WITH MULTICHANNEL INTEGRATING DETECTORS [J].
BILHORN, RB ;
EPPERSON, PM ;
SWEEDLER, JV ;
DENTON, MB .
APPLIED SPECTROSCOPY, 1987, 41 (07) :1125-1136
[8]  
BILOHORN RB, 1990, P SPIE RAMAN LUMINES, V1336, P159
[9]   A NEW STATIONARY HADAMARD ENCODING MASK FOR NEAR-INFRARED HADAMARD-TRANSFORM RAMAN-SPECTROMETRY [J].
BOHLKE, AP ;
JARVIS, JM ;
WHITE, JS ;
TATE, JD ;
PAUKSTELIS, JV ;
HAMMAKER, RM ;
FATELEY, WG .
JOURNAL OF MOLECULAR STRUCTURE, 1991, 247 :293-303
[10]   APPLICATIONS OF LASER RAMAN-SPECTROMETRY IN PROCESS-CONTROL, USING OPTICAL FIBERS [J].
GANTNER, E ;
STEINERT, D .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 338 (01) :2-8