DETERMINATION OF THE COLLECTION EFFICIENCY OF A MICROBIAL AIR SAMPLER

被引:14
作者
BENBOUGH, JE [1 ]
BENNETT, AM [1 ]
PARKS, SR [1 ]
机构
[1] CTR APPL MICROBIOL & RES,PUBL HLTH LAB SERV,DIV BIOL,BIOSAFETY UNIT,PORTON DOWN SP4 0JG,WILTS,ENGLAND
来源
JOURNAL OF APPLIED BACTERIOLOGY | 1993年 / 74卷 / 02期
关键词
D O I
10.1111/j.1365-2672.1993.tb03011.x
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The Biotest RCS Plus air sampler was compared with the well-established Casella slit sampler for the quantitative analysis of the microbiological content of air. The results of the tests reported here show that the RCS Plus sampling efficiency is similar to that of the Casella sampler over the range of particles likely to be encountered in the environment. For particles less than 4 mum down to the sub-micronic sizes the efficiency of sampling falls off only gradually so that the efficiency of sampling for 1.0 mum particles is only reduced to about 50%. The ability to pre-select 10 different volumes (from 1 to 1000 1) enables the samplers to be used for measuring a wide range of concentrations of airborne micro-organisms in a variety of locations.
引用
收藏
页码:170 / 173
页数:4
相关论文
共 8 条
[1]  
BENNETT AM, 1990, SEPARATIONS BIOTECHN, V2, P592
[2]   A slit samiler for collecting and counting air-borne bacteria [J].
Bourdillon, RB ;
Lidwell, OM ;
Thomas, JC .
JOURNAL OF HYGIENE, 1941, 41 (02) :197-224
[3]  
Clark S, 1981, J Hosp Infect, V2, P181, DOI 10.1016/0195-6701(81)90028-1
[4]  
DARK FA, 1972, SAFETY MICROBIOLOGY
[5]   AIRBORNE INFECTION IN A FULLY AIR-CONDITIONED HOSPITAL .2. TRANSFER OF AIRBORNE PARTICLES BETWEEN ROOMS RESULTING FROM MOVEMENT OF AIR FROM ONE ROOM TO ANOTHER [J].
FOORD, N ;
LIDWELL, OM .
JOURNAL OF HYGIENE, 1975, 75 (01) :31-44
[6]   THE CASCADE IMPACTOR - AN INSTRUMENT FOR SAMPLING COARSE AEROSOLS [J].
MAY, KR .
JOURNAL OF SCIENTIFIC INSTRUMENTS, 1945, 22 (10) :187-195
[7]  
MAY KR, 1973, AEROSOL SCI, V4, P235
[8]  
1988, OFFICIAL J EUROPEAN