FTIR microscopic imaging of carcinoma tissue section with 4x and 15x objectives: Practical considerations

被引:4
作者
Beleites, Claudia [1 ]
Guntinas-Lichius, O. [2 ]
Ernst, G. [2 ]
Popp, Juergen [1 ,3 ,4 ]
Krafft, Christoph [1 ]
机构
[1] Leibniz Inst Photon Technol, D-07745 Jena, Germany
[2] Univ Hosp Jena, Dept Otorhinolaryngol, Jena, Germany
[3] Univ Jena, Inst Phys Chem, D-07743 Jena, Germany
[4] Univ Jena, Abbe Ctr Photon, Jena, Germany
关键词
Infrared spectroscopy; infrared microscopy; tissue diagnosis; chemometric data analysis;
D O I
10.3233/BSI-140101
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The advent of FTIR microscopic spectrometers with focal plane array detectors enabled rapid image acquisition with diffraction limited lateral resolution. The field of view depends on the magnification and the detector size. FTIR images of large samples can be collected in the so called mosaic mode by stitching individual images together. If the mosaic is composed of hundreds of images, the total acquisition time and the data size will increase considerably. One computational and two optical options are compared to reduce both acquisition time and data size. First, the 2x field expansion optic increases the measured sample area fourfold. Second, using a 4x objective instead of the standard 15x objective increases the area covered by a single image by a factor of 11. Third, pixel binning averages neighboring pixels at the expense of lateral resolution. All options are demonstrated in a case study of a thin section of laryngeal carcinoma encompassing normal tissue, inflammation, connective tissue, dysplasia, carcinoma and blood. Data analysis is described using the toolbox hyperSpec operating under the R environment and complemented by parallel computing functions. A classification model that was trained with low magnification data in the range from 1200 to 1800 cm(-1) gave similar results for higher magnification data. Restrictions occurred for microscopic features smaller than the detector pixel size and for biomarkers below 1200 cm(-1) due to signal attenuation of the 4x objective lenses. FTIR imaging mosaic strategies of other groups and the emerging use of quantum cascade lasers for IR imaging are discussed.
引用
收藏
页码:57 / 66
页数:10
相关论文
共 22 条
[1]   Large scale infrared imaging of tissue micro arrays (TMAs) using a tunable Quantum Cascade Laser (QCL) based microscope [J].
Bassan, Paul ;
Weida, Miles J. ;
Rowlette, Jeremy ;
Gardner, Peter .
ANALYST, 2014, 139 (16) :3856-3859
[2]   Whole organ cross-section chemical imaging using label-free mega-mosaic FTIR microscopy [J].
Bassan, Paul ;
Sachdeva, Ashwin ;
Shanks, Jonathan H. ;
Brown, Mick D. ;
Clarke, Noel W. ;
Gardner, Peter .
ANALYST, 2013, 138 (23) :7066-7069
[3]  
Beleites C., 2012, HYPERSPEC PACKAGE HA
[4]  
Beleites C., 2012, CBMODELS VERSION 0 5
[6]  
DIEM M, 2008, VIBRATIONAL SPECTROS
[7]   Molecular pathology via IR and Raman spectral imaging [J].
Diem, Max ;
Mazur, Antonella ;
Lenau, Kathleen ;
Schubert, Jen ;
Bird, Ben ;
Miljkovic, Milos ;
Krafft, Christoph ;
Popp, Juergen .
JOURNAL OF BIOPHOTONICS, 2013, 6 (11-12) :855-886
[8]  
Fuller S., 2012, OPENBLASTHREADS MINI
[9]  
Griffiths PR., 2007, FOURIER TRANSFORM IN
[10]   Immunohistochemistry, histopathology and infrared spectral histopathology of colon cancer tissue sections [J].
Kallenbach-Thieltges, Angela ;
Grosserueschkamp, Frederik ;
Mosig, Axel ;
Diem, Max ;
Tannapfel, Andrea ;
Gerwert, Klaus .
JOURNAL OF BIOPHOTONICS, 2013, 6 (01) :88-100