DETERMINATION OF SERUM URIC-ACID BY ISOTOPE-DILUTION MASS-SPECTROMETRY AS A NEW CANDIDATE DEFINITIVE METHOD

被引:43
作者
ELLERBE, P
COHEN, A
WELCH, MJ
WHITE, E
机构
[1] Gaithersburg, Maryland 20899, Gaithersburg
关键词
D O I
10.1021/ac00219a004
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new Isotope dilution mass spectrometric method for uric acid Is described. A known weight of [1,3-15N2]uric acid is added to a known weight of serum, and the mixture is allowed to equilibrate. The serum is put through an anion-exchange resin, and the isolated uric acid Is converted to the tetrakis-(tert -butyldlmethylsilyl) derivative of uric acid. For measurement, the derivative is injected into a gas Chromatograph Interfaced with a low-resolution, magnetic sector mass spectrometer. Isotope ratio measurements are made from the abundances of the [M - tert-butyl]+ ions am/z 567 and 569. Blas is investigated by measuring the uric acid level In the same samples under different chromatographic conditions and with different ionization techniques. If these confirmatory measurements agree with the principal measurements, we have strong evidence for the absence of measurement bias. Uric acid was determined In three lyophlllzed human serum pools by this method. For Standard Reference Material (SRM) 909, four sets of six samples each were prepared. For Candidate SRM 909a, which consisted of two pools, each with a different level of uric acid, six sets of two samples of each level were prepared. The coefficient of variation for a single measurement ranged from 0.34% to 0.42%, while the relative standard error of the mean ranged from 0.08% to 0.14%. The results from the confirmatory measurements demonstrated that there was no significant bias in the measurements. The combination of high precision and absence of significant bias in the results qualifies this method as a candidate definitive method as defined by the National Committee for Clinical Laboratory Standards. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:2173 / 2177
页数:5
相关论文
共 19 条
  • [1] COHEN A, 1980, CLIN CHEM, V26, P854
  • [2] COHEN A, 1979, 27TH ANN C MASS SPEC
  • [3] ELLERBE P, 1988, CLIN CHEM, V34, P2280
  • [4] DETERMINATION OF SERUM-CHOLESTEROL BY A MODIFICATION OF THE ISOTOPE-DILUTION MASS-SPECTROMETRIC DEFINITIVE METHOD
    ELLERBE, P
    MEISELMAN, S
    SNIEGOSKI, LT
    WELCH, MJ
    WHITE, E
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (15) : 1718 - 1723
  • [5] CANDIDATE DEFINITIVE METHOD FOR THE DETERMINATION OF CORTISOL IN HUMAN-SERUM
    JONCKHEERE, JA
    DELEENHEER, AP
    [J]. BIOMEDICAL MASS SPECTROMETRY, 1983, 10 (03): : 197 - 202
  • [6] SYSTEMATIC-ERRORS IN GAS CHROMATOGRAPHY MASS SPECTROMETRY ISOTOPE RATIO MEASUREMENTS
    MATTHEWS, DE
    HAYES, JM
    [J]. ANALYTICAL CHEMISTRY, 1976, 48 (09) : 1375 - 1382
  • [7] PATTERSON DG, 1984, CLIN CHEM, V30, P619
  • [8] PELLETIER O, 1987, CLIN CHEM, V33, P1397
  • [9] PELLETIER O, 1987, CLIN CHEM, V33, P1403
  • [10] INHIBITION BY SOMATOSTATIN OF GASTRIN RELEASE AND GASTRIC-ACID RESPONSES TO MEALS AND TO PENTAGASTRIN IN MAN
    PHILLIP, J
    DOMSCHKE, S
    DOMSCHKE, W
    URBACH, HJ
    REISS, M
    DEMLING, L
    [J]. SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1977, 12 (03) : 261 - 265