DEFECTS IN AN INJECTION-MOLDED INTRAOCULAR-LENS

被引:0
作者
VILLARREAL, FR [1 ]
ALFONSO, EC [1 ]
ALBERT, DM [1 ]
FORSTER, RK [1 ]
机构
[1] HARVARD UNIV,SCH MED,MASSACHUSETTS EYE & EAR INFIRM,DEPT OPHTHALMOL,HOWE LAB OPHTHALMOL,BOSTON,MA 02115
关键词
CATARACT SURGERY; INJECTION MOLDED; INTRAOCULAR LENS; POSTMORTEM EXAMINATION; TRIPLE PROCEDURE;
D O I
10.1016/S0886-3350(13)80318-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Injection-molded intraocular lenses may undergo structural changes when exposed to solvents and fixatives during post-mortem examination. We report a case in which this phenomenon was seen in an injection-molded lens. We were able to duplicate these changes by immersing injection-molded lenses in 60% alcohol for 48 hours.
引用
收藏
页码:417 / 419
页数:3
相关论文
共 3 条
[1]  
Apple DJ, Craythorn JM, Olson RJ, Et al., Anterior segment complications and neovascular glaucoma following implantation of a posterior chamber intraocular lens, Ophthalmology, 91, pp. 403-419, (1984)
[2]  
Williams DJ, Polymer Science and Engineering, pp. 210-211, (1971)
[3]  
Ford ESP, Intraocular lens implantation design, Intraocular Lens Implantation, pp. 74-89, (1984)