IMPROVED DOPPLER ULTRASONIC MONITORING OF THE FETAL HEART-RATE

被引:4
作者
TUCK, DL
机构
[1] Electron. Group, Auckland Industr. Developm. Div., Dept. Sci. Industr. Res., Auckland, New Zealand
关键词
FETAL MONITORING - HEART RATE - NONINVASIVE MEASUREMENTS - ULTRASONIC CW DOPPLER TECHNIQUES;
D O I
10.1007/BF02442706
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The fetal heart rate information obtained by this technique is considered to be less precise than that obtained from the fetal ECG. The waveforms recorded by the Doppler ultrasonic method are complex and variable and this often leads to difficulty in identifying a reliable and regular measuring point within these waveforms from which to calculate the heart rate. This has resulted in a cautious acceptance of ultrasonically determined fetal heart rate records. This paper does not dispute the difficulties of interpreting such rate information, but outlines a pair of algorithms for processing and presenting a more accurate and reliable indication of the fetal heart rate determined from the Doppler ultrasonic method. These algorithms calculate a two second average rate from only those intervals which have been determined to be 'valid', based on the criterion that they should be within plus or minus 10% of the most previous valid interval.
引用
收藏
页码:135 / 140
页数:6
相关论文
共 5 条
[1]  
HON EH, 1975, INTRO FOETAL HEART R
[2]  
MANTELL CD, 1976, ATLAS FOETAL MONITOR
[3]  
OKEEFE DF, 1979, CONT OB GYN, V14, P25
[4]  
Shenker L, 1966, Obstet Gynecol Surv, V21, P367
[5]  
Wells PNT, 1972, ULTRASONICS CLIN DIA