NONINVASIVE ELECTRICAL CHARACTERIZATION OF MATERIALS AT MICROWAVE-FREQUENCIES USING AN OPEN-ENDED COAXIAL LINE - TEST OF AN IMPROVED CALIBRATION TECHNIQUE

被引:161
作者
MISRA, D
CHABBRA, M
EPSTEIN, BR
MIROTZNIK, M
FOSTER, KR
机构
[1] DAVID SARNOFF RES CTR INC,PRINCETON,NJ 08540
[2] UNIV PENN,DEPT BIOENGN,PHILADELPHIA,PA 19104
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
D O I
10.1109/22.44150
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We consider dielectric measurements using a probe consisting of coaxial transmission line with an open-circuit end placed against the sample. For the 2.99 or 3.6 mm (OD) probes considered in this study, a simple lumped parameter model shows errors above 1 GHz that increase greatly with frequency. We evaluate an approximate model due originally to Marcuvitz, on the basis of measured probe impedances from 1 to 18 GHz with samples consisting of water, methanol, and dioxane-water mixtures. This model is more accurate than the lumped-parameter model, and is better suited for calibration of the automated network analyzer. Finally, we consider the errors introduced in dielectric measurements by the use of approximate models for the probe. The technique succeeds because of partial cancellation of errors in modeling the probe in ANA-based measurements. © 1990 IEEE
引用
收藏
页码:8 / 14
页数:7
相关论文
共 16 条
[1]   MEASUREMENT OF RADIO-FREQUENCY PERMITTIVITY OF BIOLOGICAL TISSUES WITH AN OPEN-ENDED COAXIAL LINE .1. [J].
ATHEY, TW ;
STUCHLY, MA ;
STUCHLY, SS .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1982, 30 (01) :82-86
[2]   INVIVO PROBE MEASUREMENT TECHNIQUE FOR DETERMINING DIELECTRIC-PROPERTIES AT VHF THROUGH MICROWAVE-FREQUENCIES [J].
BURDETTE, EC ;
CAIN, FL ;
SEALS, J .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1980, 28 (04) :414-427
[3]   DIELECTRIC CONSTANT FOR THE DIOXANE WATER SYSTEM FROM 20 TO 35-DEGREES [J].
CRITCHFIELD, FE ;
GIBSON, JA ;
HALL, JL .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1953, 75 (08) :1991-1992
[4]   RESONANCES IN THE DIELECTRIC ABSORPTION OF DNA [J].
FOSTER, KR ;
EPSTEIN, BR ;
GEALT, MA .
BIOPHYSICAL JOURNAL, 1987, 52 (03) :421-425
[5]   THE DIELECTRIC DISPERSION OF DIOXAN-WATER MIXTURES [J].
HASTED, JB ;
HAGGIS, GH ;
HUTTON, P .
TRANSACTIONS OF THE FARADAY SOCIETY, 1951, 47 (06) :577-580
[6]   ON THE MEASUREMENT ACCURACY OF THE TISSUE PERMITTIVITY INVIVO [J].
KRASZEWSKI, A ;
STUCHLY, SS ;
STUCHLY, MA ;
SYMONS, SA .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1983, 32 (01) :37-42
[7]   ANA CALIBRATION METHOD FOR MEASUREMENTS OF DIELECTRIC-PROPERTIES [J].
KRASZEWSKI, A ;
STUCHLY, MA ;
STUCHLY, SS .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1983, 32 (02) :385-387
[8]   THEORY OF THE CIRCULAR DIFFRACTION ANTENNA [J].
LEVINE, H ;
PAPAS, CH .
JOURNAL OF APPLIED PHYSICS, 1951, 22 (01) :29-43
[9]  
Marcuvitz N., 1951, WAVEGUIDE HDB, P213
[10]   DIELECTRIC MEASUREMENTS WITH AN OPEN-ENDED COAXIAL PROBE [J].
MARSLAND, TP ;
EVANS, S .
IEE PROCEEDINGS-H MICROWAVES ANTENNAS AND PROPAGATION, 1987, 134 (04) :341-349