MANAGEMENT OF RADIATION ONCOLOGY PATIENTS WITH IMPLANTED CARDIAC-PACEMAKERS - REPORT OF AAPM TASK GROUP NO-34

被引:170
作者
MARBACH, JR
SONTAG, MR
VANDYK, J
WOLBARST, AB
机构
[1] DUKE UNIV, MED CTR, DURHAM, NC 27710 USA
[2] ONTARIO CANC INST, TORONTO M4X 1K9, ON, CANADA
[3] US EPA, WASHINGTON, DC 20460 USA
关键词
RADIATION ONCOLOGY; PACEMAKERS; TREATMENT PROTOCOL; COMPLICATIONS;
D O I
10.1118/1.597259
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Contemporary cardiac pacemakers can fail from radiation damage at doses as low as 10 gray and can exhibit functional changes at doses as low as 2 gray. A review and discussion of this potential problem is presented and a protocol is offered that suggests that radiation therapy patients with implanted pacemakers be planned so as to limit accumulated dose to the pacemaker to 2 gray. Although certain levels and types of electromagnetic interference can cause pacemaker malfunction, there is evidence that this is not a serious problem around most contemporary radiation therapy equipment.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 90 条
[41]  
HARDAGE M, 1990, COMMUNICATION
[42]   IRRADIATION OF PULMONARY TUMOR WITH OVERLYING ARTIFICIAL CARDIAC PACEMAKER [J].
HILDNER, FJ ;
LINHART, JW ;
POOLE, DO .
RADIOLOGY, 1969, 92 (01) :148-&
[43]  
HUGHES GW, 1979, SOLID STATE TECHNOL
[44]  
HUGHES GW, PRRL76CR3 RCA LABS R
[45]  
HUGHES HL, 1975, Patent No. 38629305
[46]   MEASUREMENTS OF DOSE FROM SECONDARY RADIATION OUTSIDE A TREATMENT FIELD [J].
KASE, KR ;
SVENSSON, GK ;
WOLBARST, AB ;
MARKS, MA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (08) :1177-1183
[47]   PACEMAKER FAILURE DUE TO RADIATION-THERAPY [J].
KATZENBERG, CA ;
MARCUS, FI ;
HEUSINKVELD, RS ;
MAMMANA, RB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :156-159
[48]  
KING EE, 1975, IEEE T NUCL SCI, V22, P5
[49]  
KJAR RA, 1975, C74454501 ROCKW INTC
[50]  
KJAR RA, 1974, UNPUB JUL IEEE ANN C, P203