FRESH GAS REQUIREMENTS OF AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM IN ANESTHETIZED, SPONTANEOUSLY VENTILATING CHILDREN
被引:7
|作者:
MEAKIN, G
[1
]
JENNINGS, AD
[1
]
BEATTY, PCW
[1
]
HEALY, TEJ
[1
]
机构:
[1] UNIV MANCHESTER,WITHINGTON HOSP,DEPT ANAESTHESIA,MANCHESTER M20 8LR,ENGLAND
关键词:
ANESTHESIA;
PEDIATRIC;
EQUIPMENT;
BREATHING SYSTEMS;
EAR SYSTEM;
D O I:
10.1093/bja/68.4.333
中图分类号:
R614 [麻醉学];
学科分类号:
100217 ;
摘要:
We have determined the minimum fresh gas flow rate (VF) for use with the Ohmeda enclosed afferent reservoir breathing system (EAR) in 10 anaesthetized children breathing spontaneously. First, we determined the VF required to prevent rebreathing as detected by increased total ventilation (VE) and end-tidal carbon dioxide partial pressure. Second, we used a mathematical model to calculate the degree of rebreathing occurring at each VF. A VF equal to the predicted alveolar ventilation was sufficient to prevent clinically detectable rebreathing in all patients. From the model, no rebreathing occurred when VF/VE was 0.78 or more. We have shown previously that the EAR functions efficiently during controlled ventilation with a VF = 0.6 x weight0.5. As this VF is slightly greater than the predicted alveolar ventilation, we suggest that the EAR may be used with a VF = 0.6 x weight0.5 regardless of the mode of ventilation.
引用
收藏
页码:333 / 337
页数:5