FRESH GAS REQUIREMENTS OF AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM IN ANESTHETIZED, SPONTANEOUSLY VENTILATING CHILDREN

被引:7
|
作者
MEAKIN, G [1 ]
JENNINGS, AD [1 ]
BEATTY, PCW [1 ]
HEALY, TEJ [1 ]
机构
[1] UNIV MANCHESTER,WITHINGTON HOSP,DEPT ANAESTHESIA,MANCHESTER M20 8LR,ENGLAND
关键词
ANESTHESIA; PEDIATRIC; EQUIPMENT; BREATHING SYSTEMS; EAR SYSTEM;
D O I
10.1093/bja/68.4.333
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have determined the minimum fresh gas flow rate (VF) for use with the Ohmeda enclosed afferent reservoir breathing system (EAR) in 10 anaesthetized children breathing spontaneously. First, we determined the VF required to prevent rebreathing as detected by increased total ventilation (VE) and end-tidal carbon dioxide partial pressure. Second, we used a mathematical model to calculate the degree of rebreathing occurring at each VF. A VF equal to the predicted alveolar ventilation was sufficient to prevent clinically detectable rebreathing in all patients. From the model, no rebreathing occurred when VF/VE was 0.78 or more. We have shown previously that the EAR functions efficiently during controlled ventilation with a VF = 0.6 x weight0.5. As this VF is slightly greater than the predicted alveolar ventilation, we suggest that the EAR may be used with a VF = 0.6 x weight0.5 regardless of the mode of ventilation.
引用
收藏
页码:333 / 337
页数:5
相关论文
共 36 条
  • [1] FRESH GAS REQUIREMENTS OF AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM IN ANESTHETIZED, SPONTANEOUSLY BREATHING ADULTS
    BARRIE, JR
    BEATTY, PCW
    CAMPBELL, IT
    MEAKIN, G
    HEALY, TEJ
    BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (04) : 468 - 470
  • [2] FRESH GAS REQUIREMENTS OF AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM DURING CONTROLLED VENTILATION IN CHILDREN
    MEAKIN, G
    JENNINGS, AD
    BEATTY, PCW
    HEALY, TEJ
    BRITISH JOURNAL OF ANAESTHESIA, 1992, 68 (01) : 43 - 47
  • [3] FRESH GAS-FLOW REQUIREMENTS OF A NEW ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM DURING SPONTANEOUS VENTILATION
    DROPPERT, PM
    MEAKIN, G
    BEATTY, PCW
    BEECH, M
    SHELLY, MP
    MORTIMER, AJ
    HEALY, TEJ
    BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (04) : P591 - P591
  • [4] DETERMINATION OF THE ONSET OF REBREATHING IN AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM IN ANESTHETIZED, SPONTANEOUSLY BREATHING ADULTS - A COMPARISON OF 3 METHODS
    BARRIE, JR
    BEATTY, PCW
    CAMPBELL, IT
    HEALY, TEJ
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 1994, 11 (03) : 187 - 191
  • [5] PHYSICAL CHARACTERISTICS OF AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM
    JENNINGS, AD
    MICHELL, BC
    BEATTY, PCW
    MEAKIN, G
    HEALY, TEJ
    BRITISH JOURNAL OF ANAESTHESIA, 1992, 68 (06) : 625 - 629
  • [6] EFFICIENCY OF AN ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM DURING CONTROLLED VENTILATION
    DROPPERT, PM
    MEAKIN, G
    BEATTY, PCW
    MORTIMER, AJ
    HEALY, TEJ
    BRITISH JOURNAL OF ANAESTHESIA, 1991, 66 (06) : 638 - 642
  • [7] EFFICIENCY OF A NEW ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEM DURING CONTROLLED VENTILATION
    DROPPERT, PM
    MEAKIN, G
    MORTIMER, AJ
    BEATTY, PCW
    HEALY, TEJ
    BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (02) : P277 - P277
  • [8] ASSESSMENT OF THE FRESH GAS-FLOW REQUIREMENTS OF THE ENCLOSED MAGILL ANESTHETIC BREATHING SYSTEM DURING SPONTANEOUS VENTILATION
    BRUCE, W
    SONI, N
    HANSORA, P
    BRITISH JOURNAL OF ANAESTHESIA, 1988, 60 (03) : P341 - P341
  • [9] AN ENCLOSED EFFERENT AFFERENT RESERVOIR SYSTEM - THE MAXIMA
    MILLER, DM
    ANAESTHESIA AND INTENSIVE CARE, 1995, 23 (03) : 284 - 291
  • [10] ENCLOSED AFFERENT RESERVOIR BREATHING SYSTEMS - DESCRIPTION AND CLINICAL-EVALUATION
    MILLER, DM
    MILLER, JC
    BRITISH JOURNAL OF ANAESTHESIA, 1988, 60 (04) : 469 - 475