RELATION BETWEEN SHOCK-RELATED MYOCARDIAL INJURY AND DEFIBRILLATION EFFICACY OF MONOPHASIC AND BIPHASIC SHOCKS IN A CANINE MODEL

被引:70
作者
OSSWALD, S [1 ]
TROUTON, TG [1 ]
ONUNAIN, SS [1 ]
HOLDEN, HB [1 ]
RUSKIN, JN [1 ]
GARAN, H [1 ]
机构
[1] HARVARD UNIV,MASSACHUSETTS GEN HOSP,SCH MED,CARDIAC UNIT,BOSTON,MA 02114
关键词
SHOCK; MYOCARDIUM; DEFIBRILLATION;
D O I
10.1161/01.CIR.90.5.2501
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Certain biphasic waveforms with specific time ratios of positive and negative components require less energy for successful defibrillation of the fibrillating ventricles than monophasic waveforms. However, if more efficient waveforms were also to be associated with more injurious effects on myocardial function, they might not provide a true biological advantage. This study investigates the relation between defibrillation efficacy and potential toxicity of monophasic and asymmetric, single capacitor, biphasic waveforms with equal durations of positive and negative components. Methods and Results The myocardial lactate extraction rate (LER) was used to measure the injurious effects on myocardial oxidative metabolism of two synchronized 35-J shocks in sinus rhythm. LER, mean arterial pressure (MAP) and, in a subset of experiments, cardiac output (CO) and coronary blood flow (CBF) were measured at baseline, 30 seconds, 60 seconds, 90 seconds, 150 seconds, 300 seconds, and 600 seconds after the shocks. In 12 dogs, three different waveforms (M 10: monophasic 10 milliseconds; BI 10: biphasic 10 milliseconds; BI 20: biphasic 20 milliseconds) were tested as series of two consecutive shocks (60 seconds apart) resulting in a total of 36 sets of data. At baseline, LER was 25+/-11%. After monophasic shocks, LER decreased significantly more than after biphasic shocks (LER at 150 seconds: M 10: -6+/-31% versus BI 10: 21+/-15% versus BI 20: 21+/-16%; M 10 versus BI 10 and M 10 versus BI 20, P<.05) and showed also a slower recovery (LER at 300 seconds: M 10: 1+/-24% versus BI 10: 20+/-11% versus BI 20: 20+/-15%; M 10 versus BI 10 and M 10 versus BI 20, P<.05). The maximal decrease in LER was 41+/-27% for M 10 compared with 18+/-15% for BI 10 and 15+/-11% for BI 20 (both, M 10 versus BI 10 and M 10 versus BI 20, P<.05). There was a similar decrease in CO and MAP, with the lowest MAP after monophasic shocks. The maximal decrease in MAP was significantly greater after M 10 compared with BI 20 (-29+/-15 mm Hg versus -13+/-11 mm Hg, P<.05). The defibrillation threshold was 18.6+/-8 J for M 10 compared with 11.5+/-4.0 J for BI 10 (P<.05) and 15.0+/-6.1 J for BI 20, respectively (P=NS). Conclusions Our results suggest that these specific biphasic waveforms are associated with less injurious effects on myocardial oxidative metabolism and hemodynamic performance. Given their higher defibrillation efficacy as well, biphasic waveforms may provide important long-term benefits in patients receiving frequent shocks from implantable cardioverter-defibrillators.
引用
收藏
页码:2501 / 2509
页数:9
相关论文
共 41 条
  • [1] ELECTRICAL STIMULATION OF METABOLISM OF HOMOGENATES AND PARTICULATES
    ABOOD, LG
    GERARD, RW
    OCHS, S
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1952, 171 (01): : 134 - 139
  • [2] A SIMPLIFIED, SINGLE-LEAD UNIPOLAR TRANSVENOUS CARDIOVERSION-DEFIBRILLATION SYSTEM
    BARDY, GH
    JOHNSON, G
    POOLE, JE
    DOLACK, GL
    KUDENCHUK, PJ
    KELSO, D
    MITCHELL, R
    MEHRA, R
    HOFER, B
    [J]. CIRCULATION, 1993, 88 (02) : 543 - 547
  • [3] IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS
    BARDY, GH
    HOFER, B
    JOHNSON, G
    KUDENCHUK, PJ
    POOLE, JE
    DOLACK, GL
    GLEVA, M
    MITCHELL, R
    KELSO, D
    [J]. CIRCULATION, 1993, 87 (04) : 1152 - 1168
  • [4] AN EFFECTIVE AND ADAPTABLE TRANSVENOUS DEFIBRILLATION SYSTEM USING THE CORONARY SINUS IN HUMANS
    BARDY, GH
    ALLEN, MD
    MEHRA, R
    JOHNSON, G
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (04) : 887 - 895
  • [5] ELECTRODE SYSTEM INFLUENCE ON BIPHASIC WAVE-FORM DEFIBRILLATION EFFICACY IN HUMANS
    BARDY, GH
    TROUTMAN, C
    JOHNSON, G
    MEHRA, R
    POOLE, JE
    DOLACK, GL
    KUDENCHUK, PJ
    GARTMAN, DM
    [J]. CIRCULATION, 1991, 84 (02) : 665 - 671
  • [6] COMPARATIVE EFFICACY OF MONOPHASIC AND BIPHASIC TRUNCATED EXPONENTIAL SHOCKS FOR NONTHORACOTOMY INTERNAL DEFIBRILLATION IN DOGS
    CHAPMAN, PD
    VETTER, JW
    SOUZA, JJ
    TROUP, PJ
    WETHERBEE, JN
    HOFFMANN, RG
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (03) : 739 - 745
  • [7] MYOCARDIAL NECROSIS FROM DIRECT-CURRENT COUNTERSHOCK - EFFECT OF PADDLE ELECTRODE SIZE AND TIME INTERVAL BETWEEN DISCHARGES
    DAHL, CF
    EWY, GA
    WARNER, ED
    THOMAS, ED
    [J]. CIRCULATION, 1974, 50 (05) : 956 - 961
  • [8] IMPROVED DEFIBRILLATION THRESHOLDS WITH LARGE CONTOURED EPICARDIAL ELECTRODES AND BIPHASIC WAVE-FORMS
    DIXON, EG
    TANG, ASL
    WOLF, PD
    MEADOR, JT
    FINE, MJ
    CALFEE, RV
    IDEKER, RE
    [J]. CIRCULATION, 1987, 76 (05) : 1176 - 1184
  • [9] IMPROVED INTERNAL DEFIBRILLATION EFFICACY WITH A BIPHASIC WAVEFORM
    FAIN, ES
    SWEENEY, MB
    FRANZ, MR
    [J]. AMERICAN HEART JOURNAL, 1989, 117 (02) : 358 - 364
  • [10] SUPERIORITY OF BIPHASIC SHOCKS IN THE DEFIBRILLATION OF DOGS BY EPICARDIAL PATCHES AND CATHETER ELECTRODES
    FLAKER, GC
    SCHUDER, JC
    MCDANIEL, WC
    STOECKLE, H
    DBEIS, M
    [J]. AMERICAN HEART JOURNAL, 1989, 118 (02) : 288 - 291