A MRI GRADIENT WAVE-FORM MODEL FOR AUTOMATED SEQUENCE CALIBRATION

被引:5
作者
BARKER, BR
ARCHER, BT
ERDMAN, WA
PESHOCK, RM
机构
[1] Department of Radiology, Division of MRI Research, University of Texas Southwestern Medical Center, Dallas
关键词
MAGNETIC RESONANCE IMAGING (MRI); GRADIENT; CALIBRATION; GRADIENT SWITCHING; COMPUTER SIMULATION;
D O I
10.1118/1.596928
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In order to set up magnetic resonance imaging (MRI) procedures of arbitrary voxel dimensions, slice orientation, and sequence timing in a reasonable time, some form of automatic gradient pulse calibration is required. One such method, involving simulation of gradient waveforms, is presented. Waveforms are modeled based on measurements of the step response. The model used divides each transition into three time regions: a "start" region in the first 0.3 ms, a "slew" region, and a "tail" region representing decay of the eddy current compensation error. In the "slew" region, the time derivative of the gradient, G'(t), is expressed as a function of G(t). The first two regions are nonlinear with respect to demand. The mean error in the simulated gradient is generally less than 0.04 mT m-1 in spin echo sequences. Image signal/noise ratios resulting from sequences calibrated using the model are within 5% of those of empirically calibrated sequences.
引用
收藏
页码:1483 / 1489
页数:7
相关论文
共 9 条
[1]   TEMPORAL AND SPATIAL-ANALYSIS OF FIELDS GENERATED BY EDDY CURRENTS IN SUPERCONDUCTING MAGNETS - OPTIMIZATION OF CORRECTIONS AND QUANTITATIVE CHARACTERIZATION OF MAGNET GRADIENT SYSTEMS [J].
BOESCH, C ;
GRUETTER, R ;
MARTIN, E .
MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) :268-284
[2]  
CHANG H, 1991, COMMUNICATION
[3]  
CHWIALKOWSKI M, 1991, COMMUNICATION
[4]   REDUCTION OF PULSED GRADIENT SETTLING TIME IN THE SUPERCONDUCTING MAGNET OF A MAGNETIC-RESONANCE INSTRUMENT [J].
JENSEN, DJ ;
BREY, WW ;
DELAYRE, JL ;
NARAYANA, PA .
MEDICAL PHYSICS, 1987, 14 (05) :859-862
[5]   EXACT TEMPORAL EDDY-CURRENT COMPENSATION IN MAGNETIC-RESONANCE IMAGING-SYSTEMS [J].
MORICH, MA ;
LAMPMAN, DA ;
DANNELS, WR ;
GOLDIE, FTD .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (03) :247-254
[6]  
SU MY, 1991, SMRM, V2, P836
[7]   OPTIMIZATION OF EDDY-CURRENT COMPENSATION [J].
VANVAALS, JJ ;
BERGMAN, AH .
JOURNAL OF MAGNETIC RESONANCE, 1990, 90 (01) :52-70
[8]   INTERACTIVE DESIGN OF MOTION-COMPENSATED GRADIENT WAVE-FORMS WITH A PERSONAL-COMPUTER SPREADSHEET PROGRAM [J].
WENDT, RE .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (01) :87-92
[9]   THE EFFECTS OF SIMULTANEOUS PULSING IN DIFFERENT GRADIENT COILS ON THE NUCLEAR-MAGNETIC-RESONANCE IMAGING OF OBLIQUE SLICES [J].
ZUR, Y ;
BENDEL, P .
MEDICAL PHYSICS, 1987, 14 (02) :172-177