FLUORESCENT-BASED LIFETIME MEASUREMENT THERMOMETER FOR USE AT SUBROOM TEMPERATURES (200-300 K)

被引:26
作者
ANGHEL, F [1 ]
ILIESCU, C [1 ]
GRATTAN, KTV [1 ]
PALMER, AW [1 ]
ZHANG, ZY [1 ]
机构
[1] CITY UNIV LONDON, DEPT ELECT ELECTR & INFORMAT ENGN, LONDON EC1V 0HB, ENGLAND
关键词
D O I
10.1063/1.1145596
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The development, construction, and evaluation of a simple ruby-based fiber optic thermometer, for use over the region 200-300 K, are described. The results are compared with those obtained from a simple theoretical model, which may be used for system calibration, and with those of previously published work, and good agreement is seen. The application of such a sensor to the food processing industry and environmental monitoring is indicated. © 1995 American Institute of Physics.
引用
收藏
页码:2611 / 2614
页数:4
相关论文
共 10 条
[1]   TEMPERATURE DEPENDENCES OF CR+3 RADIATIVE AND NONRADIATIVE-TRANSITIONS IN RUBY AND EMERALD [J].
FONGER, WH ;
STRUCK, CW .
PHYSICAL REVIEW B, 1975, 11 (09) :3251-3260
[2]  
Grattan K.T.V., 1995, FIBER OPTIC FLUORESC
[3]   DEVELOPMENT OF A HIGH-TEMPERATURE FIBEROPTIC THERMOMETER PROBE USING FLUORESCENT DECAY [J].
GRATTAN, KTV ;
PALMER, AW ;
ZHANG, Z .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (05) :1210-1213
[4]   RUBY DECAY-TIME FLUORESCENCE THERMOMETER IN A FIBER-OPTIC CONFIGURATION [J].
GRATTAN, KTV ;
SELLI, RK ;
PALMER, AW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (08) :1328-1335
[5]   RELATION BETWEEN ABSORPTION AND EMISSION IN REGION OF R LINES OF RUBY [J].
NELSON, DF ;
STURGE, MD .
PHYSICAL REVIEW, 1965, 137 (4A) :1117-&
[6]  
OVREN C, 1983, P INT C OPTICAL TECH
[7]  
WIEKERSHEIM K, 1985, P DIGITECH 85 BOSTON, P887
[8]  
ZHANG ZY, 1993, P SOC PHOTO-OPT INS, V1885, P300, DOI 10.1117/12.144719
[9]   FIBER OPTIC TEMPERATURE SENSOR BASED ON THE CROSS REFERENCING BETWEEN BLACKBODY RADIATION AND FLUORESCENCE LIFETIME [J].
ZHANG, ZY ;
GRATTAN, KTV ;
PALMER, AW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (05) :3177-3181
[10]   TEMPERATURE DEPENDENCES OF FLUORESCENCE LIFETIMES IN CR-3+-DOPED INSULATING CRYSTALS [J].
ZHANG, ZY ;
GRATTAN, KTV ;
PALMER, AW .
PHYSICAL REVIEW B, 1993, 48 (11) :7772-7778