Patient Setup Aid with Wireless CCTV System in Radiation Therapy

被引:0
作者
Park, Yang Kyun [1 ]
Ha, Sung Whan [2 ,3 ]
Ye, Sung Joan [2 ]
Cho, Woong [1 ]
Park, Jong Min [1 ]
Park, Suk Won [4 ]
Huh, Soon Nyung [3 ,5 ]
机构
[1] Seoul Natl Univ, Grad Sch, Dept Radiat Applying Life Sci, Seoul, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiat Oncol, Seoul, South Korea
[3] Seoul Natl Univ, Inst Radiat Med, Med Res Ctr, Seoul, South Korea
[4] Chung Ang Univ, Coll Med, Dept Radiat Oncol, Seoul, South Korea
[5] Seoul Natl Univ Hosp, Dept Radiat Oncol, Seoul, South Korea
来源
RADIATION ONCOLOGY JOURNAL | 2006年 / 24卷 / 04期
关键词
IGRT; Patient setup; Wireless CCTV; Image processing; Stereo matching; Respiratory gating;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop a wireless CCTV system in semi-beam's eye view (BEV) to monitor daily patient setup in radiation therapy. Materials and Methods: In order to get patient images in semi-BEV. CCTV cameras are installed in a custom-made acrylic applicator below the treatment head of a linear accelerator. The images from the cameras are transmitted via radio frequency signal (-2.4 GHz and 10 mW RF output). An expected problem with this system is radio frequency interference, which is solved utilizing RF shielding with Cu foils and median filtering software. The images are analyzed by our custom-made software. In the software, three anatomical landmarks in the patient surface are indicated by a user, then automatically the 3 dimensional structures are obtained and registered by utilizing a localization procedure consisting mainly of stereo matching algorithm and Gauss Newton optimization. This algorithm is applied to phantom images to investigate the setup accuracy. Respiratory gating system is also researched with real-time image processing. A line-laser marker projected on a patient's surface is extracted by binary image processing and the breath pattern is calculated and displayed in real-time. Results: More than 80% of the camera noises from the linear accelerator are eliminated by wrapping the camera with copper foils. The accuracy of the localization procedure is found to be on the order of 1.5+0.7 mm with a point phantom and sub p-millimeters and degrees with a custom-made head/neck phantom. With line -laser marker, real-time respiratory monitoring is possible in the delay time of-0.17 sec. Conclusion: The wireless CCTV camera system is the novel tool which can monitor daily patient setups. The feasibility of respiratory gating system with the wireless CCTV is hopeful.
引用
收藏
页码:300 / 308
页数:9
相关论文
共 15 条
  • [1] Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients
    Bert, C
    Metheany, KG
    Doppke, KP
    Taghian, AG
    Powell, SN
    Chen, GTY
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (04): : 1265 - 1274
  • [2] A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup
    Bert, C
    Metheany, KG
    Doppke, K
    Chen, GTY
    [J]. MEDICAL PHYSICS, 2005, 32 (09) : 2753 - 2762
  • [3] COWLEY TG, 1999, IEEE CAN C EL COMP E, P1516
  • [4] CRANE R, 1997, SIMPLE APPROACH IMAG, P95
  • [5] FristonKJ Ashburner J., 2002, HUMAN BRAIN FUNCTION, P12
  • [6] Lee S, 2003, J KOREAN SOC THER RA, V21, P100
  • [7] LI S, 2005, INTENSITY MODULATED, P407
  • [8] Patient positioning using detailed three-dimensional surface data for patients undergoing conformal radiation therapy for carcinoma of the prostate: A feasibility study
    MacKay, RI
    Graham, PA
    Logue, JP
    Moore, CJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (01): : 225 - 230
  • [9] X-ray volumetric imaging in image-guided radiotherapy: The new standard in on-treatment imaging
    McBain, CA
    Henry, AM
    Sykes, J
    Amer, A
    Marchant, T
    Moore, CM
    Davies, J
    Stratford, J
    McCarthy, C
    Porritt, B
    Williams, P
    Khoo, VS
    Price, P
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (02): : 625 - 634
  • [10] Design of 4D treatment planning target volumes
    Rietzel, Eike
    Liu, Arthur K.
    Doppke, Karen P.
    Wolfgang, John A.
    Chen, Aileen B.
    Chen, George T. Y.
    Choi, Noah C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (01): : 287 - 295