RIGHT AND LEFT-VENTRICULAR STROKE VOLUME MEASUREMENTS WITH VELOCITY-ENCODED CINE MR IMAGING - INVITRO AND INVIVO VALIDATION

被引:203
作者
KONDO, C
CAPUTO, GR
SEMELKA, R
FOSTER, E
SHIMAKAWA, A
HIGGINS, CB
机构
[1] UNIV CALIF SAN FRANCISCO, MED CTR, DEPT RADIOL, SAN FRANCISCO, CA 94143 USA
[2] UNIV CALIF SAN FRANCISCO, MED CTR, DEPT CARDIOL, SAN FRANCISCO, CA 94143 USA
[3] GE, MED SYST, MILWAUKEE, WI 53201 USA
关键词
D O I
10.2214/ajr.157.1.2048544
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The accuracy of measurements of flow velocity determined by using cine MR phase velocity mapping-velocity-encoded cine (VEC) MR-was assessed by comparing VEC MR data with independent measurements in a flow phantom and in human subjects. Constant flow velocities generated in a phantom (range, 20-408 cm/sec) were determined correctly by VEC MR (r = .997, standard error of the estimate [SEE] = 7.9 cm/sec). Peak systolic velocities in the main pulmonary artery determined by VEC MR correlated well with the measurements obtained by using continuous-wave Doppler echocardiography (r = .91). Stroke volumes measured at the aorta by VEC MR and continuous-wave Doppler imaging also correlated well with each other (r = .80). VEC MR measurements of aortic and pulmonary flow provided left and right ventricular stroke volumes that correlated well with left ventricular stroke volumes determined by short-axis cine MR images (r = .98, SEE = 3.7 ml, and r = .95, SEE = 4.8 ml, respectively). Intra- and interobserver variabilities were small for both left and right ventricular stroke volumes as measured with VEC MR. These results indicate that VEC MR accurately and reproducibly measures aortic and pulmonary flow velocities and volumes in the physiologic range of humans, and can be used to measure right and left ventricular stroke volumes under normal flow conditions.
引用
收藏
页码:9 / 16
页数:8
相关论文
共 32 条
[1]  
AXEL L, 1986, Magnetic Resonance Imaging, V4, P199, DOI 10.1016/0730-725X(86)91059-3
[2]   BLOOD-FLOW EFFECTS IN MAGNETIC-RESONANCE IMAGING [J].
AXEL, L .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) :1157-1166
[3]  
BORGEN HG, 1989, AM HEART J, V117, P1214
[4]   MEASUREMENT OF LEFT-VENTRICULAR STROKE VOLUME USING CONTINUOUS WAVE DOPPLER ECHOCARDIOGRAPHY OF THE ASCENDING AORTA AND M-MODE ECHOCARDIOGRAPHY OF THE AORTIC-VALVE [J].
BOUCHARD, A ;
BLUMLEIN, S ;
SCHILLER, NB ;
SCHLITT, S ;
BYRD, BF ;
PORTS, T ;
CHATTERJEE, K .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (01) :75-83
[5]   MEASUREMENT OF FLOW WITH NMR IMAGING USING A GRADIENT PULSE AND PHASE DIFFERENCE TECHNIQUE [J].
BRYANT, DJ ;
PAYNE, JA ;
FIRMIN, DN ;
LONGMORE, DB .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (04) :588-593
[6]  
CARO CG, 1987, MECHANICS CIRCULATIO, P57
[7]   INPLANE FLOW VELOCITY QUANTIFICATION ALONG THE PHASE ENCODING AXIS IN MRI [J].
DUERK, JL ;
PATTANY, PM .
MAGNETIC RESONANCE IMAGING, 1988, 6 (03) :321-333
[8]   MAGNETIC-RESONANCE ANGIOGRAPHY [J].
DUMOULIN, CL ;
HART, HR .
RADIOLOGY, 1986, 161 (03) :717-720
[9]  
DUMOULIN CL, 1988, MED MAGNETIC RESONAN
[10]   QUANTIFICATION OF BLOOD-FLOW WITH DYNAMIC MR IMAGING AND PRESATURATION BOLUS TRACKING [J].
EDELMAN, RR ;
MATTLE, HP ;
KLEEFIELD, J ;
SILVER, MS .
RADIOLOGY, 1989, 171 (02) :551-556