HIGH-PRECISION GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY AT LOW SIGNAL LEVELS

被引:27
作者
GOODMAN, KJ [1 ]
BRENNA, JT [1 ]
机构
[1] CORNELL UNIV,DIV NUTR SCI,ITHACA,NY 14853
关键词
D O I
10.1016/0021-9673(94)00969-G
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Precision and accuracy of gas chromatography-combustion isotope ratio mass spectrometry are investigated for sample levels down to about 5 pmol C in fatty acid methyl ester mixtures spanning 1000-fold in concentration. Precision and accuracy of isotope ratios diverge rapidly for conventional summation methods, and become unusable below 30 pmol material on column. At lower levels, mean isotope ratios were statistically different from reference values indicating bias as well as poor precision. In contrast, curve fitting, using the exponentially modified Gaussian line shape, gives improved precision for most peaks and useful results down to 3 pmol. The curve-fitting algorithm was also less sensitive to signal integration time than the summation method. These data indicate that curve fitting may be the method of choice for integration of noisy data when high-precision isotope ratios are desired.
引用
收藏
页码:63 / 68
页数:6
相关论文
共 18 条
[1]   COMPUTER ANALYSIS OF UNRESOLVED NON-GAUSSIAN GAS CHROMATOGRAMS BY CURVE-FITTING [J].
ANDERSON, AH ;
GIBB, TC ;
LITTLEWOOD, AB .
ANALYTICAL CHEMISTRY, 1970, 42 (04) :434-+
[2]   COMPUTER RESOLUTION OF UNRESOLVED CONVOLUTED GAS-CHROMATOGRAPHIC PEAKS [J].
ANDERSON, AH ;
GIBB, TC ;
LITTLEWO.AB .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1970, 8 (11) :640-&
[3]  
BRENNA JT, 1994, IN PRESS ACC CHEM RE
[4]   HIGH-PRECISION LIQUID CHROMATOGRAPHY-COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY [J].
CAIMI, RJ ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1993, 65 (23) :3497-3500
[5]  
CAIMI RJ, 1995, IN PRESS J MASS SPEC
[6]  
Cochran W.G, 1957, STAT METHODS, V6th ed
[7]  
Dyson N., 1990, CHROMATOGRAPHIC INTE
[8]   HIGH-SENSITIVITY TRACER DETECTION USING HIGH-PRECISION GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY AND HIGHLY ENRICHED [U-C-13]-LABELED PRECURSORS [J].
GOODMAN, KJ ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1088-1095
[9]   CURVE-FITTING FOR RESTORATION OF ACCURACY FOR OVERLAPPING PEAKS IN GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY [J].
GOODMAN, KJ ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1994, 66 (08) :1294-1301
[10]  
GOODMAN KJ, 1993, 41ST ANN ASMS C MASS, pA738