INTRAOPERATIVE BLOOD-GLUCOSE MEASUREMENTS - THE EFFECT OF HEMATOCRIT ON GLUCOSE TEST STRIPS

被引:19
作者
SMITH, EA [1 ]
KILPATRICK, ES [1 ]
机构
[1] WESTERN INFIRM & ASSOCIATED HOSP,DEPT PATHOL BIOCHEM,GLASGOW G11 6NT,SCOTLAND
关键词
BLOOD GLUCOSE; MEASUREMENT; REAGENT STRIPS; HEMATOCRIT;
D O I
10.1111/j.1365-2044.1994.tb03369.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Variations in haematocrit are known to affect the accuracy of reagent strip tests for glucose. We have investigated 10 patients during cardiopulmonary bypass, where intra‐operative decreases in haematocrit occur. Whole blood glucose concentrations were measured on five occasions at 30 min intervals during the procedure using the Glucometer II, One Touch II and Reflolux II meters as well as a reference instrument (YSI Model 23AM). Haematocrits were recorded simultaneously. Overall, for every 10% fall in haematocrit, Glucometer II measurements rose by 22% (r = 0.74, p < 0.00001), One Touch II measurements fell by 3% (r = 0.44, p < 0.002) and the Reflolux II measurements showed no significant variation. The One Touch II showed closer agreement to the reference (mean bias 0.3 mmol.l‐1 (95% between +0.86 and –0.26)) than the Reflolux II (bias 1.58 (+3.40 to –0.24)) or the Glucometer II (bias 3.25 (+6.18 to 0.32)). Thus, depending on the meter used, spuriously large intraoperative changes in blood glucose may seem to arise where patient haematocrit varies. Copyright © 1994, Wiley Blackwell. All rights reserved
引用
收藏
页码:129 / 132
页数:4
相关论文
共 13 条
[1]   EFFECT OF HEMATOCRIT CONCENTRATION ON BLOOD-GLUCOSE VALUE DETERMINED ON GLUCOMETER-II [J].
BARREAU, PB ;
BUTTERY, JE .
DIABETES CARE, 1988, 11 (02) :116-118
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   ARE BLOOD-GLUCOSE REAGENT STRIPS RELIABLE IN RENAL-FAILURE [J].
CLARK, JDA ;
GOLDBERG, L ;
JONES, K ;
HARTOG, M .
DIABETIC MEDICINE, 1991, 8 (02) :168-171
[4]   MORBIDITY AND MORTALITY IN DIABETICS IN FRAMINGHAM POPULATION - 16-YEAR FOLLOW-UP STUDY [J].
GARCIA, MJ ;
MCNAMARA, PM ;
GORDON, T ;
KANNELL, WB .
DIABETES, 1974, 23 (02) :105-111
[5]   THE EFFECT OF VARIATIONS IN HEMATOCRIT, MEAN CELL-VOLUME AND RED-BLOOD-CELL COUNT ON REAGENT STRIP TESTS FOR GLUCOSE [J].
KILPATRICK, ES ;
RUMLEY, AG ;
MYINT, H ;
DOMINICZAK, MH ;
SMALL, M .
ANNALS OF CLINICAL BIOCHEMISTRY, 1993, 30 :485-487
[6]  
KILPATRICK ES, 1993, P ASS CLIN BIOCH NAT
[7]   PATTERNS OF REFERRAL AND RECOVERY IN WOMEN AND MEN UNDERGOING CORONARY-ARTERY BYPASS-GRAFTING [J].
KING, KB ;
CLARK, PC ;
HICKS, GL .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (03) :179-182
[8]  
STEINER G, 1981, DIABETES, V30, P1
[9]   THE EFFECT OF HEMATOCRIT ON REAGENT STRIP TESTS FOR GLUCOSE [J].
WIENER, K .
DIABETIC MEDICINE, 1991, 8 (02) :172-175
[10]   AN ASSESSMENT OF THE EFFECT OF HEMATOCRIT ON THE HEMOCUE BLOOD-GLUCOSE ANALYZER [J].
WIENER, K .
ANNALS OF CLINICAL BIOCHEMISTRY, 1993, 30 :90-93