Radiation Dose Measurements in a 256-Slice Computed Tomography Scanner

被引:5
作者
Weir, Victor J. [1 ]
Zhang, Jie [2 ]
机构
[1] Baylor Scott & White Healthcare Syst, Med Phys & Radiat Safety, Dallas, TX 75246 USA
[2] Univ Kentucky, Div Med Phys, Lexington, KY 40536 USA
关键词
Computed tomography dose index for GE revolution; cone-beam computed tomography dose; dose profile integral; wide-beam computed tomography dosimetry;
D O I
10.4103/jmp.JMP_129_17
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to compare computed tomography (CT) radiation dose measurement methods proposed by TG111, International Electrotechnical Commission (IEC), and a direct dose profile integral (DPI) measurement method. Methods: Pencil and Farmer ion chambers are used for integrating dose profiles at different beam widths in a 60 cm long body phantom. Resulting DPI is used to calculate CT dose index (CTDI) at each beam width. Measurements are also done for a pencil chamber inserted into a 15 cm body phantom at the reference beam width. The reference measurement is scaled with pencil chamber measurements in air at different beam widths, according to the IEC approach. Finally, point dose measurements are done with a Farmer chamber under equilibrium conditions according to the TG111 method. All CTDIs calculated from measured data are compared to the scanner displayed CTDIs. Results: Calculated CTDIs, at different beam widths, using the IEC approach are within 20% of CTDIs calculated from DPI measurements in a 60 cm long body phantom. Dose Length Integral (DLI) obtained from TG111 method is close to the results obtained from DPI measurements. Scanner displayed CTDIs are lower than all measured values by up to 38% at the techniques used. Conclusion: Although the IEC method is the easiest to use compared to the TG111 and direct DPI measurement method, it underestimates dose indices by about 20%. CTDIs displayed on the GE scanner are lower than those measured in this study by up to 38%.
引用
收藏
页码:85 / 92
页数:8
相关论文
共 17 条
[1]  
[Anonymous], 2010, 111 TG AAPM
[2]   An improved analytical model for CT dose simulation with a new look at the theory of CT dose [J].
Dixon, RL ;
Munley, MT ;
Bayram, E .
MEDICAL PHYSICS, 2005, 32 (12) :3712-3728
[3]  
Dixon RL, 2003, MED PHYS, V30, P1272, DOI 10.1118/1.1576952
[4]   Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber:: Beyond CTDI100 [J].
Dixon, Robert L. ;
Ballard, Adam C. .
MEDICAL PHYSICS, 2007, 34 (08) :3399-3413
[5]   Dose equations for tube current modulation in CT scanning and the interpretation of the associated CTDIvol [J].
Dixon, Robert L. ;
Boone, John M. .
MEDICAL PHYSICS, 2013, 40 (11)
[6]   Cone beam CT dosimetry: A unified and self-consistent approach including all scan modalities-With or without phantom motion [J].
Dixon, Robert L. ;
Boone, John M. .
MEDICAL PHYSICS, 2010, 37 (06) :2703-2718
[7]   Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner [J].
Geleijns, J. ;
Artells, M. Salvado ;
Bruin, P. Wde ;
Matter, R. ;
Muramatsu, Y. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (10) :3141-3159
[8]  
Hsieh J, 2019, COMPUT TOMOGR
[9]  
IEC, 2016, MED EL EQ 2
[10]   Pseudohelical Scan for the Dose Profile Measurements of 160-mm-Wide Cone-Beam MDCT [J].
Lin, Pei-Jan P. ;
Herrnsdorf, Lars .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (04) :897-902