A KNOWLEDGE-BASED ALARM SYSTEM FOR MONITORING CARDIAC OPERATED PATIENTS - ASSESSMENT OF CLINICAL-PERFORMANCE

被引:21
作者
KOSKI, EMJ
SUKUVAARA, T
MAKIVIRTA, A
KARI, A
机构
[1] Department of Intensive Care, Kuopio University Hospital, Kuopio
[2] Medical Engineering Laboratory, NTT Information Technology Technical Research Centre of Finland, Tampere
来源
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING | 1994年 / 11卷 / 02期
关键词
KNOWLEDGE BASED SYSTEM; EVALUATION; VALIDATION; DATA FUSION; INTELLIGENT ALARMS; PATIENT MONITORING; INTENSIVE CARE;
D O I
10.1007/BF01259556
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
An intelligent alarm system for the postoperative monitoring of cardiac surgery patients, which did not require any manual data entries, was tested in two phases. A clinician monitored at bedside the patients' recovery and verified clinically abnormal physiological states. After the first test with ten patients, the system's rulebase was upgraded and then tested with an additional 15 patients. The alarm system employed two PC/ATs and was programmed to give notice of four pathological states (hyperdynamic state, hypovolemic state, hypoventilation and left ventricular failure) at two levels of urgency (alarm and alert levels). The monitoring lasted 5.4 +/- 1.7 hours per patient (mean +/- S.D.), totalling 134.7 hours. The system alarmed 27 times during the first and 73 times during the second phase of the testing. The sensitivity of the alarms was 100% in both phases, and the specificities increased from 20.0% to 73.9% and from 59.1% to 70.0% for the alarms and the alerts, respectively. This computerized decision support system based exclusively on data available in the automatically collected data base had a low false positive rate and gave early warnings about pathological states in the homogeneous group of adult postoperative cardiac patients.
引用
收藏
页码:79 / 83
页数:5
相关论文
共 16 条
[1]  
Bradshaw K.E., Gardner R.M., Clemmer T.P., Orme J.F., Thomas F., West B.J., Physician decision-making-Evaluation of data used in a computerized ICU, Int J Clin Monit Comp, 1, pp. 81-91, (1984)
[2]  
Kari A., Ruokonen E., Takala J., Comparison of acceptance and performance of automated and manual data management systems in intensive care, Int J Clin Monit Comp, 7, pp. 157-62, (1990)
[3]  
Tuman K.J., McCarthy R.J., Spiess B.D., DaValle M., Dabir R., Ivankovich A.D., Does choice of anesthetic agent significantly affect outcome after coronary artery surgery?, Anesthesiology, 70, pp. 189-98, (1989)
[4]  
Philip J.H., Thoughtful alarms, The automated anesthesia record and alarm systems, pp. 191-201, (1987)
[5]  
Koski E.M.J., Makivirta A., Sukuvaara T., Kari A., Frequency and reliability of alarms in the monitoring of cardiac postoperative patients, Int J Clin Mon Comp, 7, pp. 129-33, (1990)
[6]  
Loeb R.G., Brunner J.X., Westenskow D.R., Feldman B., Pace N.L., The Utah anesthesia workstation, Anesthesiology, 70, pp. 999-1007, (1989)
[7]  
Brunner J.X., Westenskow D.R., Zelenkov P., Prototype ventilator and alarm algorithm for NASA space station, Int J Clin Monit Comp, 5, pp. 90-9, (1989)
[8]  
Shepard L.C., Kouchoukos L.T., Kurtts M.A., Kirklin J.W., Automated treatment of critically ill patients following operation, Ann Surg, 168, (1968)
[9]  
Stoodley K.D.C., Crew A.D., Lu R., Naghdy F., A microcomputer implementation of status and alarm algorithms in a cardiac surgical intensive care unit, Int J Clin Mon Comp, 4, pp. 115-22, (1987)
[10]  
Cohn A.I., Rosenbaum S., Factor M., Miller P.L., DYNASCENE: an approach to computer-based intelligent cardiovascular monitoring using sequential clinical ‘scenes’, Meth Inf Med, 29, pp. 122-31, (1990)