Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010

被引:13
|
作者
Bramlage, Peter [1 ]
Deutsch, Cornelia [1 ]
Krueger, Ralf [1 ]
Wolf, Andreas [2 ]
Mueller, Peter [2 ]
Zwingers, Thomas [1 ,4 ]
Beime, Beate [1 ]
Mengden, Thomas [3 ]
机构
[1] Inst Pharmakol & Pravent Med, Bahnhofstrasse 20, D-49661 Cloppenburg, Germany
[2] Muller & Sebastiani, Ottobrunn, Germany
[3] Kerckhoff Klin, Bad Nauheim, Germany
[4] Estimate, Augsburg, Germany
关键词
validation; ambulatory blood pressure monitoring; ESH;
D O I
10.2147/VHRM.S63602
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: The aim of the present study was to validate the custo screen 400 ambulatory blood pressure-monitoring (ABPM) device according to the 2010 International Protocol revision of the European Society of Hypertension (ESH-IP). The device can be used for ABPM for up to 72 hours. Materials and methods: Systolic and diastolic blood pressure (SBP and DBP, respectively) were sequentially measured in 33 adult subjects (13 males and 20 females) and compared with a standard mercury sphygmomanometer (two observers). A total of 99 comparison pairs were obtained. Results: The custo screen 400 met the requirements of parts 1 and 2 of the ESH-IP revision 2010. The mean difference between the device and reference sphygmomanometer readings was -0.5 +/- 4.5 mmHg for SBP and -0.1 +/- 3.3 mmHg for DBP. All but one measurement were within the absolute difference of 10 mmHg between the device and the observers for SBP and DBP. The number of absolute differences between the device and the observers within a range of 5 mmHg was 84 of 99 readings for SBP, and 93 of 99 readings for DBP. Conclusion: The custo screen 400 ABPM device met the requirements of the 2010 ESH-IP revision, and hence can be recommended for ABPM in adults. To our knowledge, the custo screen 400 is the first device to pass the revised ESH-IP 2010.
引用
收藏
页码:303 / 309
页数:7
相关论文
共 50 条
  • [41] Validation of the iHealth BP7 wrist blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Wang, Qing
    Zhao, Huadong
    Chen, Wan
    Li, Ni
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 54 - 57
  • [42] Validation of the Medisana MTP Plus upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Erdem, Emre
    Aydogdu, Turkan
    Akpolat, Tekin
    BLOOD PRESSURE MONITORING, 2011, 16 (01) : 43 - 46
  • [43] Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Song, Cheng
    Yu, Yang
    Lu, Bao-Chuan
    Yan, Xi-Ling
    BLOOD PRESSURE MONITORING, 2020, 25 (05) : 291 - 294
  • [44] European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults (vol 15, pg 23, 2010)
    O'Brien, E.
    Atkins, N.
    Stergiou, G.
    Karpettas, N.
    Parati, G.
    Asmar, R.
    BLOOD PRESSURE MONITORING, 2010, 15 (03) : 171 - 172
  • [45] Validation of the A&D BP UA-651 device with a wide-range cuff for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2015, 20 (03) : 164 - 167
  • [46] Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (06) : 329 - 331
  • [47] Evaluation of Meditech ABPM-06 ambulatory blood pressure measuring device, according to the European Society of Hypertension, the British Hypertension Society and the International Organization for Standardization Protocol
    Hermanyi, Zsolt
    Pokoly, Bence
    Visolyi, Gergely
    Barna, Istvan
    BLOOD PRESSURE MONITORING, 2019, 24 (04) : 208 - 211
  • [48] Validation of the Andon KD-5851 upper arm blood pressure monitor, for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Wu, Lili
    Jiao, Yinghui
    Wang, Chengdong
    Chen, Lei
    Di, Dalin
    Zhang, Haiyan
    BLOOD PRESSURE MONITORING, 2015, 20 (04) : 228 - 231
  • [49] Validation in the General Population of the iHealth Track Blood Pressure Monitor for Self-Measurement According to the European Society of Hypertension International Protocol Revision 2010: Descriptive Investigation
    Mazoteras-Pardo, Victoria
    Becerro-De-Bengoa-Vallejo, Ricardo
    Elena Losa-Iglesias, Marta
    Lopez-Lopez, Daniel
    Palomo-Lopez, Patricia
    Rodriguez-Sanzl, David
    Calvo-Lobo, Cesar
    JMIR MHEALTH AND UHEALTH, 2019, 7 (03):
  • [50] VALIDATION OF THREE AUTOMATIC BLOOD PRESSURE MONITORS FOR SELF-MEASUREMENT ACCORDING TO EUROPEAN SOCIETY OF HYPERTENSION INTERNATIONAL PROTOCOL REVERSION 2010
    Tian, H.
    Zeng, S. Y.
    Zhong, Z. Y.
    Gong, W.
    Liu, W. J.