Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010

被引:13
|
作者
Bramlage, Peter [1 ]
Deutsch, Cornelia [1 ]
Krueger, Ralf [1 ]
Wolf, Andreas [2 ]
Mueller, Peter [2 ]
Zwingers, Thomas [1 ,4 ]
Beime, Beate [1 ]
Mengden, Thomas [3 ]
机构
[1] Inst Pharmakol & Pravent Med, Bahnhofstrasse 20, D-49661 Cloppenburg, Germany
[2] Muller & Sebastiani, Ottobrunn, Germany
[3] Kerckhoff Klin, Bad Nauheim, Germany
[4] Estimate, Augsburg, Germany
关键词
validation; ambulatory blood pressure monitoring; ESH;
D O I
10.2147/VHRM.S63602
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: The aim of the present study was to validate the custo screen 400 ambulatory blood pressure-monitoring (ABPM) device according to the 2010 International Protocol revision of the European Society of Hypertension (ESH-IP). The device can be used for ABPM for up to 72 hours. Materials and methods: Systolic and diastolic blood pressure (SBP and DBP, respectively) were sequentially measured in 33 adult subjects (13 males and 20 females) and compared with a standard mercury sphygmomanometer (two observers). A total of 99 comparison pairs were obtained. Results: The custo screen 400 met the requirements of parts 1 and 2 of the ESH-IP revision 2010. The mean difference between the device and reference sphygmomanometer readings was -0.5 +/- 4.5 mmHg for SBP and -0.1 +/- 3.3 mmHg for DBP. All but one measurement were within the absolute difference of 10 mmHg between the device and the observers for SBP and DBP. The number of absolute differences between the device and the observers within a range of 5 mmHg was 84 of 99 readings for SBP, and 93 of 99 readings for DBP. Conclusion: The custo screen 400 ABPM device met the requirements of the 2010 ESH-IP revision, and hence can be recommended for ABPM in adults. To our knowledge, the custo screen 400 is the first device to pass the revised ESH-IP 2010.
引用
收藏
页码:303 / 309
页数:7
相关论文
共 50 条
  • [31] Validation of the A&D BP UB-542 wrist device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Saladini, Francesca
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2013, 18 (04) : 219 - 222
  • [32] Validation of the A&D BP UB-543 wrist device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Fania, Claudio
    Benetti, Elisabetta
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2015, 20 (04) : 237 - 240
  • [33] Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol
    Coleman, A
    Freeman, P
    Steel, S
    Shennan, A
    BLOOD PRESSURE MONITORING, 2005, 10 (03) : 165 - 168
  • [34] Validation of Transtek blood pressure monitor TMB-1491 for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Tian, Huiyong
    Zeng, Sijian
    Zhong, Xiaoyan
    Gong, Wei
    Liu, Wenjun
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 280 - 282
  • [35] A NOVEL DUAL-MODE AMBULATORY AND HOME BLOOD PRESSURE MONITOR: VALIDATION ACCORDING TO THE EUROPEAN SOCIETY OF HYPERTENSION INTERNATIONAL PROTOCOL
    Stergiou, G. S.
    Tzamouranis, D.
    Nasothimiou, E.
    Karpettas, N.
    Protogerou, A.
    JOURNAL OF HYPERTENSION, 2009, 27 : S195 - S195
  • [36] Validation of Transtek TMB-1776 according to European Society of Hypertension International Protocol revision 2010 in adults
    Liu, Zhonghua
    Chen, Liyi
    BLOOD PRESSURE MONITORING, 2019, 24 (06) : 319 - 322
  • [37] Validation of the Nissei DS-250 ambulatory blood pressure monitoring device according to the International Protocol
    Altunkan, S
    Iliman, N
    Altunkan, E
    JOURNAL OF HYPERTENSION, 2003, 21 : S22 - S22
  • [38] Validation of the Tensoval Duo Control II blood pressure monitor for clinic use and self-measurement according to the British Hypertension Society protocol and the European Society of Hypertension International Protocol revision 2010
    de Greeff, Annemarie
    Shennan, Andrew H.
    BLOOD PRESSURE MONITORING, 2013, 18 (03) : 161 - 166
  • [39] Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Huang, Qi-Fang
    Wang, Jie
    Sheng, Chang-Sheng
    Zhang, Na-Na
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (04) : 232 - 234
  • [40] Validation of the HONSUN LD-578 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zhang, Yi
    Wang, Jie
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2009, 14 (03) : 128 - 131