Validation of the custo screen 400 ambulatory blood pressure-monitoring device according to the European Society of Hypertension International Protocol revision 2010

被引:13
|
作者
Bramlage, Peter [1 ]
Deutsch, Cornelia [1 ]
Krueger, Ralf [1 ]
Wolf, Andreas [2 ]
Mueller, Peter [2 ]
Zwingers, Thomas [1 ,4 ]
Beime, Beate [1 ]
Mengden, Thomas [3 ]
机构
[1] Inst Pharmakol & Pravent Med, Bahnhofstrasse 20, D-49661 Cloppenburg, Germany
[2] Muller & Sebastiani, Ottobrunn, Germany
[3] Kerckhoff Klin, Bad Nauheim, Germany
[4] Estimate, Augsburg, Germany
关键词
validation; ambulatory blood pressure monitoring; ESH;
D O I
10.2147/VHRM.S63602
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: The aim of the present study was to validate the custo screen 400 ambulatory blood pressure-monitoring (ABPM) device according to the 2010 International Protocol revision of the European Society of Hypertension (ESH-IP). The device can be used for ABPM for up to 72 hours. Materials and methods: Systolic and diastolic blood pressure (SBP and DBP, respectively) were sequentially measured in 33 adult subjects (13 males and 20 females) and compared with a standard mercury sphygmomanometer (two observers). A total of 99 comparison pairs were obtained. Results: The custo screen 400 met the requirements of parts 1 and 2 of the ESH-IP revision 2010. The mean difference between the device and reference sphygmomanometer readings was -0.5 +/- 4.5 mmHg for SBP and -0.1 +/- 3.3 mmHg for DBP. All but one measurement were within the absolute difference of 10 mmHg between the device and the observers for SBP and DBP. The number of absolute differences between the device and the observers within a range of 5 mmHg was 84 of 99 readings for SBP, and 93 of 99 readings for DBP. Conclusion: The custo screen 400 ABPM device met the requirements of the 2010 ESH-IP revision, and hence can be recommended for ABPM in adults. To our knowledge, the custo screen 400 is the first device to pass the revised ESH-IP 2010.
引用
收藏
页码:303 / 309
页数:7
相关论文
共 50 条
  • [1] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beime, Beate
    Deutsch, Cornelia
    Krueger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Hammel, Gertrud
    Bramlage, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2017, 176 (05) : 573 - 580
  • [2] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 573 - 580
  • [3] Erratum to: Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 835 - 836
  • [4] Validation of the Hingmed WBP-02A device for ambulatory blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Fania, Claudio
    Vezzu, Leonardo
    Lazzaretto, Ilaria
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2019, 24 (03) : 151 - 154
  • [5] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010 (vol 176, pg 573, 2017)
    Beime, Beate
    Deutsch, Cornelia
    Kruger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Hammel, Gertrud
    Bramlage, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2017, 176 (06) : 835 - 836
  • [6] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    M Abou-Dakn
    C Döhmen
    S Wenzel
    Journal of Human Hypertension, 2017, 31 : 89 - 92
  • [7] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    Abou-Dakn, M.
    Doehmen, C.
    Wenzel, S.
    JOURNAL OF HUMAN HYPERTENSION, 2017, 31 (02) : 89 - 92
  • [8] Validation of the blood pressure measurement device Erkameter 125 PRO according to the European Society of Hypertension International Protocol revision 2010
    Beime, Beate
    Deutsch, Cornelia
    Krueger, Ralf
    Zimmermann, Erik
    Bramlage, Peter
    BLOOD PRESSURE MONITORING, 2016, 21 (06) : 356 - 360
  • [9] Validation of the blood pressure measurement device Beurer BM 28 according to the European Society of Hypertension International Protocol revision 2010
    Deutsch, Cornelia
    Bramlage, Carsten
    Botta, Beate
    Krueger, Ralf
    Forstner, Klaus
    Bramlage, Peter
    Beime, Beate
    BLOOD PRESSURE MONITORING, 2021, 26 (04) : 292 - 298
  • [10] Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010
    Michael Abou-Dakn
    Sabine Wenzel
    Journal of Human Hypertension, 2018, 32 : 770 - 774