DUAL X-RAY ABSORPTIOMETRY FOR THE EVALUATION OF BONE-DENSITY FROM THE PROXIMAL FEMUR AFTER TOTAL HIP-ARTHROPLASTY - ANALYSIS PROTOCOLS AND REPRODUCIBILITY

被引:60
作者
TREVISAN, C
BIGONI, M
CHERUBINI, R
STEIGER, P
RANDELLI, G
ORTOLANI, S
机构
[1] UNIV MILAN, OSPED MAGGIORE MILANO, IST SCI MED, VIA F SFORZA 35, I-20122 MILAN, ITALY
[2] IST ORTOPED G PINI, MILAN, ITALY
[3] HOLOG INC, WALTHAM, MA USA
关键词
TOTAL HIP ARTHROPLASTY; DUAL X-RAY ABSORPTIOMETRY; BONE MINERAL DENSITY;
D O I
10.1007/BF01321831
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Dual X-ray absorptiometry (DXA) instruments are now able to evaluate bone mineral density (BMD) of bone surrounding metal implants. The assessment of BMD around prosthetic components could provide additional information for the follow-up of total hip arthroplasty (THA). In this study, we evaluated the potential application of DXA in the field of THA. BMD was measured in the proximal femur of both THA and THA-free sides in 14 postmenopausal women 6-18 months after THA. The explored segment was divided into seven zones as proposed by Gruen et al. [18]. The precision error of BMD measurements ranged from 1.8 to 6.8% on the THA side and from 1.1 to 2% to the THA-free side. The reduction of BMD of the THA versus the THA-free side was significant in all seven zones (P < 0.01, t-test for paired data). These results showed significant differences in BMD around femoral components of THA with respect to contralateral healthy side, and demonstrate the sensitivity of DXA for detecting these changes.
引用
收藏
页码:158 / 161
页数:4
相关论文
共 27 条