Inferior vena cava filter misplacement in the right atrium and migration to the right ventricle followed by successful removal using the endovascular technique: A case report and review of the literature

被引:0
作者
Wakabayashi, Yasushi [1 ,2 ]
Takeuchi, Wataru [1 ]
Yamazaki, Kyohei [1 ]
机构
[1] Nagano Prefectural Kiso Hosp, Dept Cardiovasc Med, Nagano, Japan
[2] Jichi Med Univ, Saitama Med Ctr, Dept Cardiovasc Med, Saitama, Japan
关键词
Inferior vena cava filter; pulmonary embolism; deep vein thrombosis; filter migration; endovascular technique;
D O I
10.1177/2050313X15595833
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inferior vena cava filters are effective for preventing the passage of thrombi into the pulmonary arteries in patients with pulmonary embolism and deep vein thrombosis. These filters are indicated in patients with contraindications to anticoagulant therapy or in patients with recurrent acute pulmonary embolism despite the administration of anticoagulant therapy. However, the occurrence of filter-related complications, such as filter migration to the heart, has been increasing. Herein, we report a case of OptEase inferior vena cava filter misplacement in the right atrium. Although the filter migrated to the right ventricle, it was successfully removed and repositioned in the inferior vena cava using endovascular techniques. Unfortunately, moderate tricuspid regurgitation developed, due to the damage to the tricuspid valve that was caused by the procedure. We have also reviewed the relevant literature and discussed the possible strategies for managing cases of filter migration to the heart and preventing filter misplacement.
引用
收藏
页数:4
相关论文
共 9 条
[1]   Right ventricular foreign body: Percutaneous transvenous retrieval of a Greenfield filter from the right ventricle - A case report [J].
Arjomand, H ;
Surabhi, S ;
Wolf, NM .
ANGIOLOGY, 2003, 54 (01) :109-113
[2]   Right ventricular migration and endovascular removal of an inferior vena cava filter [J].
Bui, James T. ;
West, Derek L. ;
Pinto, Christina ;
Gramling-Babb, Patricia ;
Owens, Charles A. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (01) :141-144
[3]   Sudden cardiac death caused by migration of a TrapEase inferior vena cava filter: Case report and review of the literature [J].
Haddadian, Babak ;
Shaikh, Fareed ;
Djelmami-Hani, Mohamed ;
Shalev, Yoseph .
CLINICAL CARDIOLOGY, 2008, 31 (02) :84-87
[4]  
Izutani H, 2004, CAN J CARDIOL, V20, P233
[5]   Emergency retrieval of a G2 filter after complete migration in to the right ventricle [J].
Kuo, William T. ;
Loh, Christopher T. ;
Sze, Daniel Y. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 18 (09) :1177-1182
[6]   Intracardiac Migration of Inferior Vena Cava Filters Review of Published Data [J].
Owens, Charles A. ;
Bui, James T. ;
Knuttinen, M-Grace ;
Gaba, Ron C. ;
Carrillo, Tami C. ;
Hoefling, Nickoleta ;
Layden-Almer, Jennifer E. .
CHEST, 2009, 136 (03) :877-887
[7]  
Peters MN, 2013, TEX HEART I J, V40, P316
[8]   Successful percutaneous retrieval of an inferior vena cava filter migrating to the right ventricle in a bariatric patient [J].
Veerapong J. ;
Wahlgren C.M. ;
Jolly N. ;
Bassiouny H. .
CardioVascular and Interventional Radiology, 2008, 31 (Suppl 2) :S177-S181
[9]   Clinical review: Inferior vena cava filters in the age of patient-centered outcomes [J].
Wang, Stephen L. ;
Lloyd, Allen J. .
ANNALS OF MEDICINE, 2013, 45 (07) :474-481