MAGNETIC-RESONANCE, A NEW METHOD FOR MEASURING BLOOD-FLOW IN HEMODIALYSIS FISTULAS

被引:38
|
作者
OUDENHOVEN, LFIJ [1 ]
PATTYNAMA, PMT [1 ]
DEROOS, A [1 ]
SEEVERENS, HJJ [1 ]
REBERGEN, SA [1 ]
CHANG, PC [1 ]
机构
[1] UNIV HOSP LEIDEN,DEPT NEPHROL,2300 RC LEIDEN,NETHERLANDS
关键词
D O I
10.1038/ki.1994.116
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Blood flow through arteriovenous fistulae in patients undergoing hemodialysis needs to be within a certain range for optimal management. Magnetic resonance (MR) velocity mapping is a new modality that allows the noninvasive measurement of blood flow volume. This technique was applied in 13 patients with either a Brescia-Cimino shunt (N = 6) or a Polytetrafluoroethylene graft (Goretex, N = 7). Fistula flow volume measured by MR velocity mapping was 1386 +/- 411 ml/min (mean +/- SD, range 644 to 1921 ml/min). Interstudy reproducibility of MR velocity mapping was good (r = 0.94). MR velocity mapping derived flow volumes showed good correlation with values obtained with an indicator dilution technique using Indocyanine Green (ICG) which was performed simultaneously in eight patients (r = 0.86). We conclude that MR velocity mapping is an accurate noninvasive method to quantify flow volume through arteriovenous fistulae in patients requiring hemodialysis.
引用
收藏
页码:884 / 889
页数:6
相关论文
共 50 条
  • [1] BLOOD-FLOW - MAGNETIC-RESONANCE IMAGING
    BRADLEY, WG
    WALUCH, V
    RADIOLOGY, 1985, 154 (02) : 443 - 450
  • [2] BLOOD-FLOW IMAGING BY MAGNETIC-RESONANCE
    WATANABE, S
    SAKAKIBARA, M
    IMAI, H
    MASUDA, Y
    INAGAKI, Y
    MORITA, F
    UEMATSU, S
    BIORHEOLOGY, 1985, 22 (04) : 359 - 359
  • [3] BLOOD-FLOW IMAGING BY CINE MAGNETIC-RESONANCE
    NAYLER, GL
    FIRMIN, DN
    LONGMORE, DB
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1986, 10 (05) : 715 - 722
  • [4] MEASUREMENT OF BLOOD-FLOW BY NUCLEAR MAGNETIC-RESONANCE
    BATTOCLETTI, JH
    POSTGRADUATE MEDICINE, 1982, 72 (05) : 43 - &
  • [5] BLOOD-FLOW EFFECTS IN MAGNETIC-RESONANCE IMAGING
    AXEL, L
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) : 1157 - 1166
  • [6] BLOOD-FLOW IMAGING BY NUCLEAR MAGNETIC-RESONANCE
    BATTOCLETTI, JH
    HALBACH, RE
    SANCES, A
    EVANS, SM
    CLINICAL RESEARCH, 1982, 30 (04): : A758 - A758
  • [7] BLOOD-FLOW MEASUREMENTS USING MAGNETIC-RESONANCE
    YOUNG, IR
    BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (692): : 798 - 798
  • [8] BLOOD-FLOW IMAGING WITH NUCLEAR MAGNETIC-RESONANCE
    SINGER, JR
    LIBOVE, J
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1977, 27 (NOV): : 165 - 166
  • [9] MAGNETIC-RESONANCE (MR) IN THE DETERMINATION OF BLOOD-FLOW SPEED
    WEIKL, A
    MULLER, E
    REINHARDT, ER
    BACHMANN, K
    ZEITSCHRIFT FUR KARDIOLOGIE, 1986, 75 : 48 - 48
  • [10] NUCLEAR MAGNETIC-RESONANCE - PRINCIPLES OF BLOOD-FLOW IMAGING
    MILLS, CM
    BRANTZAWADZKI, M
    CROOKS, LE
    KAUFMAN, L
    SHELDON, P
    NORMAN, D
    BANK, W
    NEWTON, TH
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 142 (01) : 165 - 170