INVITRO MEASUREMENT AND CHARACTERIZATION OF CURRENT-DENSITY PROFILES PRODUCED BY NONRECESSED, SIMPLE RECESSED, AND RADIALLY VARYING RECESSED STIMULATING ELECTRODES

被引:58
作者
SUESSERMAN, MF
SPELMAN, FA
RUBINSTEIN, JT
机构
[1] UNIV WASHINGTON,DEPT ELECT ENGN,SEATTLE,WA 98195
[2] UNIV WASHINGTON,CTR BIOENGN,SEATTLE,WA 98195
[3] UNIV WASHINGTON,SCH MED,SEATTLE,WA 98195
关键词
D O I
10.1109/10.81558
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Potential fields induced by nonrecessed, simple recessed, and radially varying recessed electrode designs were measured in vitro. Comparison of experimental results with theoretical analyses substantiated the experimental measurement technique and emphasized the importance of considering both nonuniform charge injection and surface electrochemistry when designing implantable stimulating electrodes. Radially varying recesses produced uniform charge injection at the electrode surface and at the aperture-tissue interface. In general, the radially varying recessed electrodes provided a combination of uniform charge injection and flexibility in design and fabrication that warrants their incorporation into all appropriate planar stimulating electrode designs.
引用
收藏
页码:401 / 408
页数:8
相关论文
共 12 条
[1]   ELECTRICAL-STIMULATION WITH PT ELECTRODES - TRACE ANALYSIS FOR DISSOLVED PLATINUM AND OTHER DISSOLVED ELECTROCHEMICAL PRODUCTS [J].
BRUMMER, SB ;
MCHARDY, J ;
TURNER, MJ .
BRAIN BEHAVIOR AND EVOLUTION, 1977, 14 (1-2) :10-22
[2]   ELECTRICAL-STIMULATION OF NERVOUS-SYSTEM - PRINCIPLE OF SAFE CHARGE INJECTION WITH NOBLE-METAL ELECTRODES [J].
BRUMMER, SB ;
TURNER, MJ .
BIOELECTROCHEMISTRY AND BIOENERGETICS, 1975, 2 (01) :13-25
[3]  
Geddes L. A., 1972, ELECTRODES MEASUREME
[4]   NEURAL PROSTHESES [J].
HAMBRECHT, FT .
ANNUAL REVIEW OF BIOPHYSICS AND BIOENGINEERING, 1979, 8 :239-267
[5]   COCHLEAR PATHOLOGY FOLLOWING CHRONIC INTRA-COCHLEAR ELECTRICAL-STIMULATION [J].
LEAKEJONES, PA ;
WALSH, SM ;
MERZENICH, MM .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1981, 90 (02) :6-8
[6]   CURRENT-DENSITY PROFILES OF SURFACE MOUNTED AND RECESSED ELECTRODES FOR NEURAL PROSTHESES [J].
RUBINSTEIN, JT ;
SPELMAN, FA ;
SOMA, M ;
SUESSERMAN, MF .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (11) :864-875
[7]  
RUBINSTEIN JT, 1987, 9TH P ANN C IEEE ENG, P913
[8]  
RUBINSTEIN JT, 1987, THESIS U WASHINGTON
[9]  
SCOBEY RP, 1981, IEEE T BIOMED ENG, V28
[10]  
Shepherd R K, 1983, Acta Otolaryngol Suppl, V399, P19