MORPHOLOGY OF POLYMETHYLMETHACRYLATE (PMMA) BONE-CEMENT - SURFACE-STRUCTURES AND CAUSES OF THEIR ORIGIN

被引:22
作者
WILLERT, HG
MUELLER, K
SEMLITSCH, M
机构
[1] TH DARMSTADT, STATE MAT TESTING INST, DEPT PLAST, D-6100 DARMSTADT, FED REP GER
[2] SULZER BROTHERS LTD, DEPT MED ENGN RES, WINTERTHUR, SWITZERLAND
关键词
D O I
10.1007/BF00383411
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study deals with the correlation between the polymerizing bone cement and the surrounding tissue. The surface structures of bone cements, polymerized in air, in tissue medium (in vitro) and in human bone during implantation were investigated and compared with the contours of the tissue of the implant bed. Basing on the dimensional differences it was differentiated between contours of 1st order and 2nd order: contours of 1st order are within the macroscopic range, contours of 2nd order within the microscopic range. The surface of bone cement polymerized in living human tissue differed essentially from samples polymerized under laboratory conditions. The differences are to be seen macroscopically in the coarse relief as well as microscopically in the shape and the connection of the superficial methylmethacrylate beads. Bone cements, polymerized in air show an ideal, even and closed surface. Bone cements, polymerized in tissue medium exhibit macroscopically some wrinkles, in the microscopic range their contours are either closed (samples prepolymerized at 22°C) or partly open and partly closed (samples prepolymerized at 24°C). The surface of bone cement implants, retrieved from human bones are characterized macroscopically by a marked wrinkled and papillary relief, microscopically by flattened beads, and most often by an irregular, rough and open surface with isolated beads giving almost the impression of a porous surface structure. © 1979 J. F. Bergmann Verlag.
引用
收藏
页码:265 / 292
页数:28
相关论文
共 38 条
[1]   SLOW CRACK GROWTH IN ACRYLIC BONE CEMENT [J].
BEAUMONT, PWR ;
YOUNG, RJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1975, 9 (05) :423-439
[2]  
BECHTEL A, 1973, CHROMATOGRAPHIA, V5, P226
[3]  
CAMERON HU, 1974, CLIN ORTHOP RELAT R, P287
[4]  
Charnley J, 1970, ACRYLIC CEMENT ORTHO, pvi
[5]  
De Wijn J.R., 1975, J BIOMED RES S, V6, P99
[6]   WORKING TIME OF BONE CEMENTS [J].
DEBRUNNER, HU ;
WETTSTEIN, A .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1975, 81 (04) :291-299
[7]   VOLUME CHANGES OF BONE CEMENTS [J].
DEBRUNNER, HU .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1975, 81 (01) :37-44
[8]  
Debrunner HU., 1976, ENG MED, V2, P294
[9]  
EGGERT A, 1974, CHIRURG, V45, P236
[10]   ALTERATIONS OF PHYSICAL-PROPERTIES OF SO-CALLED BONE CEMENTS AFTER ADMIXING OF FOREIGN INGREDIENTS [J].
GRUNERT, A ;
RITTER, G .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1974, 78 (04) :336-342