BROKEN FILTERS IN LONG-TERM FOLLOW-UP OF GUNTHER CAVA FILTERS - RESULTS AFTER 64 FILTER IMPLANTATIONS

被引:4
作者
URBAN, M [1 ]
HRUBY, W [1 ]
WINKLER, WB [1 ]
MOSSER, H [1 ]
BALDT, M [1 ]
机构
[1] KA RUDOLFSTIFTUNG,MED ABT 2,VIENNA,AUSTRIA
来源
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN | 1992年 / 156卷 / 04期
关键词
VENA CAVA FILTERS; FILTER BREAKAGE; VENA CAVA FILTERS - COMPLICATIONS; THROMBOSIS; FILTER;
D O I
10.1055/s-2008-1032898
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Follow-up of patients with inferior vena cava filters was performed every 3 months with clinical controls, plain radiographs and duplex sonography. Whereas the clinical results were excellent (no recurrence of pulmonary embolism), radiological controls showed broken anchoring legs in 14 cases and a broken filter basket in one patient. In two cases, the broken anchoring prongs had perforated the inferior vena cava wall. All patients with broken filters were clinically asymptomatic. CT showed no morphological changes in these cases. As experience proves, strict follow-up of caval filters for early detection of complications is mandatory even with well-established filter types
引用
收藏
页码:342 / 345
页数:4
相关论文
共 18 条
  • [1] PERCUTANEOUS KIMRAY-GREENFIELD FILTER PLACEMENT BY FEMORAL VEIN PUNCTURE
    DENNY, DF
    CRONAN, JJ
    DORFMAN, GS
    ESPLIN, C
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1985, 145 (04) : 827 - 829
  • [2] GREENFIELD FILTER - PERCUTANEOUS PLACEMENT IN 50 PATIENTS
    DENNY, DF
    DORFMAN, GS
    CRONAN, JJ
    GREENWOOD, LH
    MORSE, SS
    YOSELEVITZ, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 150 (02) : 427 - 429
  • [3] PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS
    DORFMAN, GS
    [J]. RADIOLOGY, 1990, 174 (03) : 987 - 992
  • [4] DUCKSOO K, 1989, RADIOLOGY, V172, P721
  • [5] GUNTHER VENA-CAVAL FILTER - RESULTS OF LONG-TERM FOLLOW-UP
    FOBBE, F
    DIETZEL, M
    KORTH, R
    FELSENBERG, D
    BENDER, S
    HAMED, M
    LAASS, C
    SORENSEN, R
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (05) : 1031 - 1034
  • [6] GREENFIELD LJ, 1973, SURGERY, V73, P599
  • [7] GREENFIELD LJ, 1988, SURGERY, V104, P706
  • [8] 1ST CLINICAL-RESULTS WITH A NEW CAVAL FILTER
    GUNTHER, RW
    SCHILD, H
    HOLLMAN, JP
    VORWERK, D
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1987, 10 (02) : 104 - 108
  • [9] HEUYN HY, 1989, RADIOLOGY, V172, P15
  • [10] JOHNSON SG, 1988, J INTERVENT RAD, V3, P33