Effectiveness of a Drill-assisted Intraosseous Catheter versus Manual Intraosseous Catheter by Resident Physicians in a Swine Model

被引:7
作者
Hafner, John W. [1 ]
Bryant, Adam [1 ]
Huang, Felix [1 ]
Swisher, Keir [1 ]
机构
[1] Univ Illinois, Coll Med Peoria, Div Emergency Med, Peoria, IL 61656 USA
关键词
Emergency medicine; Intraosseous needle; IO; Resident physician;
D O I
10.5811/westjem.2013.4.13361
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Our objective was to compare the effectiveness, speed, and complication rate of the traditional manually placed intraosseous (IO) catheter to a mechanical drill-assisted IO catheter by emergency medicine (EM) resident physicians in a training environment. Methods: Twenty-one EM residents participated in a randomized prospective crossover experiment placing 2 intraosseous needles (Cook (R) Intraosseous Needle, Cook Medical, Bloomington, IN; and EZ-IO (R) Infusion System, Vidacare, San Antonio, TX). IO needles were placed in anesthetized mixed breed swine (mass range: 25 kg to 27.2 kg). The order of IO placement and puncture location (proximal tibia or distal femur) were randomly assigned. IO placement time was recorded from skin puncture until the operator felt they had achieved successful placement. We used 3 verification criteria: aspiration of marrow blood, easy infusion of 10 mL saline mixed with methylene blue, and lack of stained soft tissue extravasation. Successful placement was defined as meeting 2 out of the 3 predetermined criteria. We surveyed participants regarding previous IO experience, device preferences, and comfort levels using multiple choice, Likert scale, and visual analog scale (VAS) questions. IO completion times, VAS, and mean Likert scales were compared using Student's t-test and success rates were compared using Fisher's exact test with p<0.05 considered significant. Results: Drill-assisted IO needle placement was faster than manually placed IO needle placement (3.66 versus 33.57 seconds; p=0.01). Success rates were 100% with the drill-assisted IO needle and 76.2% with the manual IO needle (p=0.04). The most common complication of the manual IO insertion was a bent needle (33.3% of attempts). Participants surveyed preferred the drill-assisted IO insertion more than the manual IO insertion (p<0.0001) and felt the drill-assisted IO was easier to place (p<0.0001). Conclusion: In an experimental swine model, drill-assisted IO needle placement was faster and had less failures than manual IO needle placement by inexperienced resident physicians. EM resident physician participants preferred the drill-assisted IO needle.
引用
收藏
页码:629 / 632
页数:4
相关论文
共 17 条
[1]   Intraosseous is faster and easier than umbilical venous catheterization in newborn emergency vascular access models [J].
Abe, KK ;
Blum, GT ;
Yamamoto, LG .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2000, 18 (02) :126-129
[2]   Comparison of two intraosseous infusion systems for adult emergency medical use [J].
Brenner, Thorsten ;
Bernhard, Michael ;
Helm, Matthias ;
Doll, Sara ;
Voelkl, Alfred ;
Ganion, Nicole ;
Friedmann, Claudia ;
Sikinger, Marcus ;
Knapp, Juergen ;
Martin, Eike ;
Gries, Andre .
RESUSCITATION, 2008, 78 (03) :314-319
[3]   Intraosseous infusion devices: A comparison for potential use in special operations [J].
Calkins, MD ;
Fitzgerald, G ;
Bentley, TB ;
Burris, D .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2000, 48 (06) :1068-1074
[4]  
Davidoff Jack, 2005, JEMS, V30, P20
[5]  
Dubick MA, 2000, MIL MED, V165, P552
[6]   Consecutive field trials using two different intraosseous devices [J].
Frascone, Ralph J. ;
Jensen, Joe P. ;
Kaye, Kory ;
Salzman, Joshua G. .
PREHOSPITAL EMERGENCY CARE, 2007, 11 (02) :164-171
[7]  
Gillum Lee, 2005, JEMS, V30, P24
[8]   PEDIATRIC INTRAOSSEOUS INFUSIONS - IMPACT ON VASCULAR ACCESS TIME [J].
GLAESER, PW ;
LOSEK, JD ;
NELSON, DB ;
BONADIO, WA ;
SMITH, DS ;
WALSHKELLY, C ;
HENNES, H .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1988, 6 (04) :330-332
[9]   5-YEAR EXPERIENCE IN PREHOSPITAL INTRAOSSEOUS INFUSIONS IN CHILDREN AND ADULTS [J].
GLAESER, PW ;
HELLMICH, TR ;
SZEWCZUGA, D ;
LOSEK, JD ;
SMITH, DS .
ANNALS OF EMERGENCY MEDICINE, 1993, 22 (07) :1119-1124
[10]   Intraosseous Devices: A Randomized Controlled Trial Comparing Three Intraosseous Devices [J].
Hartholt, Klaas A. ;
van Lieshout, Esther M. M. ;
Thies, Wim C. ;
Patka, Peter ;
Schipper, Inger B. .
PREHOSPITAL EMERGENCY CARE, 2010, 14 (01) :6-13