Subclinical cardiac perforation caused by a Micra leadless pacemaker

被引:10
作者
Togashi, Ikuko [1 ]
Sato, Toshiaki [2 ]
Hoshida, Kyoko [1 ]
Soejima, Kyoko [1 ]
机构
[1] Kyorin Univ, Cardiol, Sch Med, Mitaka, Tokyo, Japan
[2] Kyorin Univ, Div Adv Arrhythmia Management, Sch Med, Mitaka, Tokyo, Japan
关键词
cardiac effusion; cardiac perforation; leadless pacemaker; Micra; transcatheter pacing system;
D O I
10.1002/joa3.12052
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A subclinical cardiac perforation by a device cup of the Micra transcatheter pacing system was suspected in a 78-year-old woman. During the procedure, the device cup was placed on the septum. The contrast media was injected before device deployment and remained outside of the myocardium. Later, a cardiac computed tomography scan visualized a protruded diverticular structure on the right ventricle. The contrast material remained in a pouch within the pericardium. To ensure the device is oriented away from the border between the right ventricular septum and the free wall, right anterior oblique view should be carefully reviewed before deployment.
引用
收藏
页码:326 / 328
页数:3
相关论文
共 50 条
[31]   A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry [J].
Roberts, Paul R. ;
Clementy, Nicolas ;
Al Samadi, Faisal ;
Garweg, Christophe ;
Martinez-Sande, Jose Luis ;
Iacopino, Saverio ;
Johansen, Jens Brock ;
Prat, Xavier Vinolas ;
Kowal, Robert C. ;
Klug, Didier ;
Mont, Lluis ;
Steffel, Jan ;
Li, Shelby ;
Van Osch, Dirk ;
El-Chami, Mikhael F. .
HEART RHYTHM, 2017, 14 (09) :1375-1379
[32]   Comparison of the safety and efficacy of Nanostim and Micra transcatheter leadless pacemaker (LP) extractions: a multicenter experience [J].
Tawseef Dar ;
Krishna Akella ;
Ghulam Murtaza ;
Sharan Sharma ;
Muhammad R. Afzal ;
Rakesh Gopinathannair ;
Ralph Augostini ;
John Hummel ;
Dhanunjaya Lakkireddy .
Journal of Interventional Cardiac Electrophysiology, 2020, 57 :133-140
[33]   Cost-effectiveness of Micra™ VR leadless pacemaker in patients with bradycardia and atrial fibrillation in Australia [J].
Makino, Koji ;
Mudge, Mia ;
Hill, Michelle ;
Zaunmayr, Chelsea ;
Tilden, Dominic .
JOURNAL OF ARRHYTHMIA, 2024, 40 (06) :1481-1489
[34]   Comparison of the safety and efficacy of Nanostim and Micra transcatheter leadless pacemaker (LP) extractions: a multicenter experience [J].
Dar, Tawseef ;
Akella, Krishna ;
Murtaza, Ghulam ;
Sharma, Sharan ;
Afzal, Muhammad R. ;
Gopinathannair, Rakesh ;
Augostini, Ralph ;
Hummel, John ;
Lakkireddy, Dhanunjaya .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 57 (01) :133-140
[35]   Development and validation of a risk score for predicting pericardial effusion in patients undergoing leadless pacemaker implantation: experience with the Micra transcatheter pacemaker [J].
Piccini, Jonathan P. ;
Cunnane, Ryan ;
Steffel, Jan ;
El-Chami, Mikhael F. ;
Reynolds, Dwight ;
Roberts, Paul R. ;
Soejima, Kyoko ;
Steinwender, Clemens ;
Garweg, Christophe ;
Chinitz, Larry ;
Ellis, Christopher R. ;
Stromberg, Kurt ;
Fagan, Dedra H. ;
Mont, Lluis .
EUROPACE, 2022, 24 (07) :1119-1126
[36]   The subtle tine: Asymptomatic Micra perforation incidentally discovered during cardiac surgery [J].
Tan, Nicholas Y. ;
Madhavan, Malini ;
Greason, Kevin L. ;
Cha, Yong-Mei .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (01) :229-230
[37]   Leadless pacemaker implantation after delayed atrial lead perforation and battery depletion: a case report [J].
Zhao, Yichang ;
Su, Liping ;
Gao, Yuchen ;
Wang, Hao ;
Luan, Chao ;
Liu, Jinqiu ;
Chen, Feifei .
BMC CARDIOVASCULAR DISORDERS, 2024, 24 (01)
[38]   How to upgrade a leadless pacemaker to cardiac resynchronization therapy [J].
Lacour, Thibaud ;
Bisson, Arnaud ;
Bernard, Anne ;
Fauchier, Laurent ;
Babuty, Dominique ;
Clementy, Nicolas .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (11) :2578-2581
[39]   Towards a Leadless Wirelessly Controlled Intravenous Cardiac Pacemaker [J].
Anwar, Usama ;
Ajijola, Olujimi A. ;
Shivkumar, Kalyanam ;
Markovic, Dejan .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2022, 69 (10) :3074-3086
[40]   Pediatric Micra leadless pacemaker implantation via internal jugular and femoral veins: experience with 11 patients [J].
El-Bokl, Amr ;
Siddeek, Hani ;
Hou, Cody ;
Leslie, Alison ;
Jimenez, Erick ;
Cortez, Daniel .
FUTURE CARDIOLOGY, 2022, 18 (09) :679-686