Subclinical cardiac perforation caused by a Micra leadless pacemaker

被引:10
作者
Togashi, Ikuko [1 ]
Sato, Toshiaki [2 ]
Hoshida, Kyoko [1 ]
Soejima, Kyoko [1 ]
机构
[1] Kyorin Univ, Cardiol, Sch Med, Mitaka, Tokyo, Japan
[2] Kyorin Univ, Div Adv Arrhythmia Management, Sch Med, Mitaka, Tokyo, Japan
关键词
cardiac effusion; cardiac perforation; leadless pacemaker; Micra; transcatheter pacing system;
D O I
10.1002/joa3.12052
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A subclinical cardiac perforation by a device cup of the Micra transcatheter pacing system was suspected in a 78-year-old woman. During the procedure, the device cup was placed on the septum. The contrast media was injected before device deployment and remained outside of the myocardium. Later, a cardiac computed tomography scan visualized a protruded diverticular structure on the right ventricle. The contrast material remained in a pouch within the pericardium. To ensure the device is oriented away from the border between the right ventricular septum and the free wall, right anterior oblique view should be carefully reviewed before deployment.
引用
收藏
页码:326 / 328
页数:3
相关论文
共 50 条
[21]   Perforation of the right atrial appendage during implantation of a leadless pacemaker [J].
Date, Kazuma ;
Murata, Tomohiro ;
Mano, Akiko ;
Kawata, Mitsuhiro ;
Kyo, Shunei .
JOURNAL OF ARRHYTHMIA, 2022, 38 (01) :163-165
[22]   First Experience of Micra™ Leadless Pacemaker Implantation Outside the Indonesian Capital [J].
Tondas, Alexander E. ;
Halim, Rolando A. ;
Tandri, Fredy .
EUROPEAN HEART JOURNAL SUPPLEMENTS, 2019, 21 (0F) :F170-F170
[23]   Tricuspid leaflet flail after Micra™ leadless pacemaker implantation: a case report [J].
Pingitore, Annachiara ;
Calcagno, Simone ;
Salvador, Loris ;
Mennuni, Silvia ;
Cavarretta, Elena .
EUROPEAN HEART JOURNAL-CASE REPORTS, 2022, 6 (06)
[24]   Is day-case surgical procedure safe for MICRA leadless pacemaker implantation? [J].
Toon, Lin-Thiri ;
Elrefai, Mohammed ;
Abouelasaad, Mohamed ;
Patil, Roopa ;
Paisey, John ;
Yue, Arthur ;
Roberts, Paul .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2025, 68 (01) :117-123
[25]   Micra-AV leadless pacemaker and atrioventricular (dys)synchrony: A stepwise process [J].
Mitacchione, Gianfranco ;
Schiavone, Marco ;
Gasperetti, Alessio ;
Ruggiero, Diego ;
Denora, Marialessia ;
Viecca, Maurizio ;
Forleo, Giovanni Battista .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (10) :1738-1742
[26]   The angle of the tines before the pull and hold test predicts engagement of the tines in Micra leadless pacemaker implantation [J].
Mizukami, Akira ;
Miyakuni, Shota ;
Nakada, Ryo ;
Kobayashi, Tetsuya ;
Kawakami, Takuya ;
Takegawa, Koki ;
Arai, Hirofumi ;
Hiroki, Jiro ;
Yoshioka, Kenji ;
Otani, Hirofumi ;
Ono, Maki ;
Yamashita, Shu ;
Ueshima, Daisuke ;
Suzuki, Makoto ;
Matsumura, Akihiko ;
Goya, Masahiko ;
Sasano, Tetsuo .
JOURNAL OF ARRHYTHMIA, 2023, 39 (01) :10-17
[27]   A case of successful salvage despite right ventricular perforation during AVEIR VR leadless pacemaker implantation [J].
Matsuo, Masanori ;
Shimeno, Kenji ;
Matsumoto, Naoki ;
Abe, Yukio ;
Fukuda, Daiju .
JOURNAL OF ARRHYTHMIA, 2025, 41 (01)
[28]   Safety and Performance of the Micra VR Leadless Pacemaker in a Japanese Cohort ― Comparison With Global Studies ― [J].
Ando, Kenji ;
Inoue, Kanki ;
Harada, Tomoo ;
Shizuta, Satoshi ;
Yoshida, Yukihiko ;
Kusano, Kengo ;
Onuki, Tatsuya ;
Watari, Yuji ;
Fukui, Akio ;
Sasaki, Shingo ;
Shoda, Morio ;
Nishii, Nobuhiro ;
Shiose, Akira ;
Hosoda, Junya ;
Okai, Chie ;
Stromberg, Kurt ;
Murphy, Jeffrey ;
Holmes, Thomas R. ;
Soejima, Kyoko .
CIRCULATION JOURNAL, 2023, 87 (12) :1809-+
[29]   Visualization and appearance of artifacts of leadless pacemaker systems in cardiac MRI [J].
Edlinger, Christoph ;
Granitz, Marcel ;
Paar, Vera ;
Jung, Christian ;
Pfeil, Alexander ;
Eder, Sarah ;
Wernly, Bernhard ;
Kammler, Juergen ;
Hergan, Klaus ;
Hoppe, Uta C. ;
Steinwender, Clemens ;
Lichtenauer, Michael ;
Kypta, Alexander .
WIENER KLINISCHE WOCHENSCHRIFT, 2018, 130 (13-14) :427-435
[30]   Assessing safety of leadless pacemaker (MICRA) at various implantation sites and its impact on paced QRS in Indian population [J].
Sharma, Prafull ;
Guleria, Vivek Singh ;
Bharadwaj, Prashant ;
Datta, Rajat .
INDIAN HEART JOURNAL, 2020, 72 (05) :376-382