Many clinical and research studies in nuclear medicine require quantitation of iodine-123 (I-123) distribution for the determination of kinetics or localization. The objective of this study was to implement several reconstruction methods designed for single-photon emission computed tomography (SPECT) using I-123 and to evaluate their performance in terms of quantitative accuracy, image artifacts, and noise. The methods consisted of four attenuation and scatter compensation schemes incorporated into both the filtered backprojection/Chang (FBP) and maximum likelihood-expectation maximization (ML-EM) reconstruction algorithms. The methods were evaluated on data acquired of a phantom containing a hot sphere of I-123 activity in a lower level background I-123 distribution and nonuniform density media. For both reconstruction algorithms, nonuniform attenuation compensation combined with either scatter subtraction or Metz filtering produced images that were quantitatively accurate to within 15% of the true value. The ML-EM algorithm demonstrated quantitative accuracy comparable to FBP and smaller relative noise magnitude for all compensation schemes.
引用
收藏
页码:527 / 533
页数:7
相关论文
共 33 条
[1]   3-DIMENSIONAL RECONSTRUCTION IN NUCLEAR-MEDICINE EMISSION IMAGING [J].
BUDINGER, TF ;
GULLBERG, GT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1974, NS21 (03) :2-20
[2]  
BUDINGER TF, 1980, J NUCL MED, V21, P579
[4]  
COLEMAN RE, 1983, EMISSION COMPUTED TO, P135
[5]   MAXIMUM LIKELIHOOD FROM INCOMPLETE DATA VIA EM ALGORITHM [J].
DEMPSTER, AP ;
LAIRD, NM ;
RUBIN, DB .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-METHODOLOGICAL, 1977, 39 (01) :1-38
[6]  
DRAYER B, 1983, AM J NEURORADIOL, V4, P572
[7]   BRAIN PHANTOM - HIGH-RESOLUTION IMAGING WITH SPECT AND I-123 [J].
FLOYD, CE ;
JASZCZAK, RJ ;
GREER, KL ;
COLEMAN, RE .
RADIOLOGY, 1987, 164 (01) :279-281
[8]  
GREER K L, 1987, Journal of Nuclear Medicine Technology, V15, P53
[9]   AN ATTENUATED PROJECTOR BACKPROJECTOR FOR ITERATIVE SPECT RECONSTRUCTION [J].
GULLBERG, GT ;
HUESMAN, RH ;
MALKO, JA ;
PELC, NJ ;
BUDINGER, TF .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (08) :799-816
[10]  
HILL TC, 1982, J NUCL MED, V23, P191