LOW-LEVEL COUNTING BY LIQUID SCINTILLATION .I. TRITIUM MEASUREMENT IN HOMOGENEOUS SYSTEMS

被引:32
作者
MOGHISSI, AA
KELLEY, HL
REGNIER, JE
CARTER, MW
机构
[1] Southeastern Radiological Helath Laboratory, Montgomery, AL 36101
来源
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES | 1969年 / 20卷 / 03期
关键词
D O I
10.1016/0020-708X(69)90026-X
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The significant factors which affect low-level counting of tritium by liquid scintillation have been investigated. In a typical operating mode of the system having a background of 8·5 counts/min, one count/min above background corresponds to 380 pCi/l of water (120 T.R.). The developed system combines convenience of automatic sample counting, room temperature operation, high efficiency, high capacity, and simplicity for effective application in measuring tritium levels in environmental samples. © 1969.
引用
收藏
页码:145 / &