FARADAY CUP DOSIMETRY IN A PROTON THERAPY BEAM WITHOUT COLLIMATION

被引:33
作者
GRUSELL, E
ISACSSON, U
MONTELIUS, A
MEDIN, J
机构
[1] UNIV UPPSALA HOSP,AKAD SJUKHUSET,DEPT HOSP PHYS,S-75185 UPPSALA,SWEDEN
[2] KAROLINSKA INST,DEPT MED RADIAT PHYS,S-17176 STOCKHOLM,SWEDEN
关键词
D O I
10.1088/0031-9155/40/11/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A Faraday cup in a proton beam can give an accurate measurement of the number of protons collected by the cup. It is shown that the collection efficiency with a proper design can be close to unity. To be able to calibrate an ionization chamber from such a measurement, as is recommended in some dosimetry protocols, the energy spectrum of the proton beam must be accurately known. This is normally not the case when the lateral beam extension is defined by collimators. Therefore a method for relating an ionization chamber measurement in an uncollimated beam to the total number of protons in the beam has been developed and is described together with experimental results from calibrating an ionization chamber using this method in the therapeutic beam in Uppsala. This method is applicable to ionization chambers of any shape and the accuracy is estimated to be 1.6% (1 SD).
引用
收藏
页码:1831 / 1840
页数:10
相关论文
共 12 条
[1]   LINEARITY WITH DOSE-RATE OF LOW-RESISTIVITY P-TYPE SILICON SEMICONDUCTOR-DETECTORS [J].
GRUSELL, E ;
RIKNER, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) :785-792
[2]  
KACPEREK A, 1990, P INT HEAVY PARTICLE, V69, P53
[3]   IONIZATION-CHAMBER DOSIMETRY OF PROTON-BEAMS USING CYLINDRICAL AND PLANE-PARALLEL CHAMBERS - N-W VERSUS N-K ION-CHAMBER CALIBRATIONS [J].
MEDIN, J ;
ANDREO, P ;
GRUSELL, E ;
MATTSSON, O ;
MONTELIUS, A ;
ROOS, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (07) :1161-1176
[4]  
Nordic Association of Clinical Physics, 1980, ACTA RADIOL ONCOL, V19, P55
[5]   SOME CONSIDERATIONS REGARDING W-VALUES FOR HEAVY CHARGED-PARTICLE RADIOTHERAPY [J].
VERHEY, LJ ;
LYMAN, JT .
MEDICAL PHYSICS, 1992, 19 (01) :151-153
[6]   DETERMINATION OF ABSORBED DOSE IN A PROTON-BEAM FOR PURPOSES OF CHARGED-PARTICLE RADIATION-THERAPY [J].
VERHEY, LJ ;
KOEHLER, AM ;
MCDONALD, JC ;
GOITEIN, M ;
MA, IC ;
SCHNEIDER, RJ ;
WAGNER, M .
RADIATION RESEARCH, 1979, 79 (01) :34-54
[7]   CODE OF PRACTICE FOR CLINICAL PROTON DOSIMETRY [J].
VYNCKIER, S ;
BONNETT, DE ;
JONES, DTL .
RADIOTHERAPY AND ONCOLOGY, 1991, 20 (01) :53-63
[8]   SUPPLEMENT TO THE CODE OF PRACTICE FOR CLINICAL PROTON DOSIMETRY [J].
VYNCKIER, S ;
BONNETT, DE ;
JONES, DTL .
RADIOTHERAPY AND ONCOLOGY, 1994, 32 (02) :174-179
[9]  
1993, ICRU49 REP
[10]  
1987, ABSORBED DOSE DETERM