Midplane dose determination and verification of calculated doses in total body irradiation

被引:2
作者
Ozdemir, Ozlem [1 ]
Hoca, Sinan [2 ]
Olacak, Nezahat [2 ]
Ozkok, Serdar [2 ]
Kamer, Serra [2 ]
Anacak, Yavuz [2 ]
机构
[1] Pamukkale Univ, Tip Fak, Radyasyon Onkol Anabilim Dali, Denizli, Turkey
[2] Ege Univ, Tip Fak, Radyasyon Onkol Anabilim Dali, Izmir, Turkey
来源
TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY | 2013年 / 28卷 / 02期
关键词
Total body irradiation; in-vivo dosimetry; thermoluminescence dosimetry;
D O I
10.5505/tjoncol.2013.898
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVES To compare calculated and measured doses for different regions of anthropomorphic phantom and patients using ion chamber and thermoluminescence dosimetry (TLD) for total body irradiation. METHODS Measurements were done for lateral fields with 6 MV, gantry 82 degrees, 40x40 cm(2) field and 400 cm source-axis distance (SAD). Entrance-exit and midline doses were measured on anthropomorphic phantom by TLD and entrance-exit doses were measured by TLD and ion chamber on patients. RESULTS For anthropomorphic phantom measurements differences between calculated and measured entrance-exit doses of head, neck, shoulder, lung and thick pelvis were 0.8%, 2.7%, 26.4%, 4.4% and 4.9% and for midline doses were 1.6%, 1.6%, 6.3%, -1.4% and 7.4% respectively. For patients; TLD differences were within -4.13% ile 6.7%, -3.3% ile 3.9%, 5.1% ile 16.6%, -7.8% ile 2.4%, and 3.6% ile 7.1% respectively. For thick pelvis measurements with ion chamber differences were within %00.1-1.9. CONCLUSION Total body irradiation is being applied in limit values in our clinic.
引用
收藏
页码:67 / 74
页数:8
相关论文
共 11 条
[1]   Dose evaluation of patients receiving total-body irradiation for the pre-treatment of bone marrow transplantation [J].
Ban, N ;
Sawai, S ;
Aoki, Y ;
Nakagawa, K ;
Kusama, T .
RADIATION PROTECTION DOSIMETRY, 1997, 71 (01) :61-64
[2]  
Bomford CK, 1994, TXB RADIOTHERAPY, p[105, 465]
[3]   Thermoluminescence dosimetry applied to in vivo dose measurements for total body irradiation techniques [J].
Duch, MA ;
Ginjaume, M ;
Chakkor, H ;
Ortega, X ;
Jornet, N ;
Ribas, M .
RADIOTHERAPY AND ONCOLOGY, 1998, 47 (03) :319-324
[4]   In vivo dosimetry during external photon beam radiotherapy [J].
Essers, M ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (02) :245-259
[5]  
Galvin CM, 1986, 17 AAPM AM I PHYS, P5
[6]  
Khan FM, 2003, PHYS RAD THERAPY, P455
[7]   CLINICAL THERMOLUMINESCENCE DOSIMETRY - HOW DO EXPECTATIONS AND RESULTS COMPARE [J].
KRON, T ;
SCHNEIDER, M ;
MURRAY, A ;
MAMEGHAN, H .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :151-161
[8]  
Malicki J, 1997, MED PHYS, V22, P30
[9]   In-vivo dosimetry by diode semiconductors in combination with portal films during TBI: reporting a 5-year clinical experience [J].
Mangili, P ;
Fiorino, C ;
Rosso, A ;
Cattaneo, GM ;
Parisi, R ;
Villa, E ;
Calandrino, R .
RADIOTHERAPY AND ONCOLOGY, 1999, 52 (03) :269-276
[10]   Midplane dose determination during total body irradiation using in vivo dosimetry [J].
Ribas, M ;
Jornet, N ;
Eudaldo, T ;
Carabante, D ;
Duch, MA ;
Ginjaume, M ;
de Segura, GG ;
Sánchez-Doblado, F .
RADIOTHERAPY AND ONCOLOGY, 1998, 49 (01) :91-98