SR-82 PRODUCTION FROM METALLIC RB TARGETS AND DEVELOPMENT OF AN RB-82 GENERATOR SYSTEM

被引:19
作者
CACKETTE, MR [1 ]
RUTH, TJ [1 ]
VINCENT, JS [1 ]
机构
[1] TRIUMF,4004 WESBROOK MALL,VANCOUVER V6T 2A3,BC,CANADA
关键词
D O I
10.1016/0969-8043(93)90045-C
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A complete procedure for producing Sr-82/Rb generators is described. Sr-82 was produced by bombarding metallic rubidium targets with 60 MeV protons with beam currents as high as 80 muA. The experimentally determined cross section for a thick target (61-48 MeV) agrees with the literature value of 100 mb. Production rates of about 0.2 mCi/muA-h with a 1.8 g/cm2 target were achieved. The Sr-82 was separated from the target by dissolving the rubidium metal in n-butanol then separating the strontium isotopes on a Chelex-100 column. A SnO2 generator system was built from readily available components. Factors affecting generator design, performance and limitations included SnO2 particle size, column length and eluant flow-rate.
引用
收藏
页码:917 / 922
页数:6
相关论文
共 12 条
  • [1] PREPARATION AND EVALUATION OF A HYDROUS TIN(IV) OXIDE 82SR/82RB MEDICAL GENERATOR SYSTEM FOR CONTINUOUS ELUTION
    BRIHAYE, C
    GUILLAUME, M
    OBRIEN, HA
    RAETS, D
    DELANDSHEERE, C
    RIGO, P
    [J]. APPLIED RADIATION AND ISOTOPES, 1987, 38 (03) : 213 - 217
  • [2] CACKETTE M, 1991, Journal of Labelled Compounds and Radiopharmaceuticals, V30, P109
  • [3] GOULD KL, 1991, J NUCL MED, V32, P579
  • [4] GUILLAUME M, 1984, RADIONUCLIDE GENERAT, P185
  • [5] GUNNICK R, 1972, AUTOMATED GAMMA SPEC, V1, P1
  • [6] HOLMES RA, 1991, NATIONAL BIOMEDICAL, P14
  • [7] HOROGUCHI T, 1980, APPL RADIAT ISOTOPES, V31, P141
  • [8] MAUSNER LF, 1987, APPL RADIAT ISOTOPES, V38, P181, DOI 10.1016/0883-2889(87)90084-0
  • [9] THOMAS KE, 1987, APPL RADIAT ISOTOPES, V38, P175, DOI 10.1016/0883-2889(87)90083-9
  • [10] WATERS SL, 1987, APPL RADIAT ISOT, V8, P171