FATIGUE-CORROSION OF ENDOPROSTHESIS TITANIUM-ALLOYS

被引:4
作者
CORNET, A
MUSTER, D
JAEGER, JH
机构
[1] HOSP CIVILS,SERV STOMATOL & CHIRURG MAXILLO FACIALE,F-67005 STRASBOURG,FRANCE
[2] CTR TRAUMATOL & ORTHOPED STRASBOURG,F-67400 ILLKIRCH GRAFFENS,FRANCE
来源
BIOMATERIALS MEDICAL DEVICES AND ARTIFICIAL ORGANS | 1979年 / 7卷 / 01期
关键词
Compendex;
D O I
10.3109/10731197909119383
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Commercial total hip prostheses often show certain metallurgical faults (porosities, coarse grains, growth dendrites, carbide networks). In oràer to investigate more accurately the role played by these different parameters in prostheses failure we performed a large number of systematic corrosion, fatigue and fatigue - corrosion tests on these materials and on commercial total hip prostheses. Ultimate strengthes seem to be reached for cast cobalt alloys, whereas titanium alloys, such as Ta 6 V., present very high fatigue limit under corrosion. Thus, rotatj.ve bending fatigue - corrosion tests in biological environment provide values about 50 DaN/mm2. This value, is nevertheless appreciably higher than those obtained with stell ‘.tes and stainless steel. Titanium alloys, because of their mechanical performances, their weak Young's modulus (11000 DaN/mm2) and their relative lightness (4.5. g/cm3), which are associated with a good biocompatibility, seem very promising for permanent implants realisation. © 1979, Informa UK Ltd. All rights reserved.
引用
收藏
页码:155 / 167
页数:13
相关论文
共 14 条
[1]  
AUVINET J, 1973, MEM ETUD SCI REV MET, V70, P809
[2]  
CORNET A, 1978, MAY M EUR SOC BIOM B
[3]  
DAUBERTES C, 1958, REV METALL, V55, P1048
[4]  
DAUBERTES C, 1960, REV METALL, P337
[5]   CORROSION-RESISTANT ALLOYS IN CHLORIDE SOLUTIONS - MATERIALS FOR SURGICAL IMPLANTS [J].
HOAR, TP ;
MEARS, DC .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1966, 294 (1439) :486-&
[6]  
Kempf I, 1976, Acta Orthop Belg, V42 Suppl 1, P194
[7]  
LAING PATRICK G., 1967, J BIOMED MAT RES, V1, P135, DOI 10.1002/jbm.820010113
[8]   CREVICE CORROSION IN ORTHOPEDIC IMPLANT METALS [J].
LEVINE, DL ;
STAEHLE, RW .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1977, 11 (04) :553-561
[9]  
LOCATI L, 1955, METALLURGIA ITALIANA, V9
[10]  
LORENTZ M, 1977, 1ST EUR C EV BIOM