DETERMINATION OF A CATECHOL-O-METHYLTRANSFERASE INHIBITOR, NITECAPONE, IN HUMAN PLASMA AND URINE BY LIQUID-CHROMATOGRAPHY

被引:8
|
作者
WIKBERG, T
TASKINEN, J
机构
[1] Orion Pharmaceutica, Research Center, 02101 Espoo
关键词
CATECHOL-O-METHYLTRANSFERASE INHIBITOR; NITECAPONE; REVERSED-PHASE LIQUID CHROMATOGRAPHY; ELECTROCHEMICAL DETECTION; PLASMA; URINE;
D O I
10.1016/0731-7085(91)80238-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Methods based on reversed-phase liquid chromatography with amperometric detection have been developed for determination of nitecapone, 3-(3,4-dihydroxy-5-nitrobenzylidene)-2,4-pentanedione, a COMT inhibitor, in human plasma and urine. Nitecapone was extracted with ethyl acetate-hexane mixtures from plasma after acidification with hydrochloric acid and from urine as the tetrabutylammonium ion-pair of its diphenylborate derivative. The recoveries of both methods exceeded 70% and the relative standard deviations for within-day precision were less than 4% and 8% at 50 ng ml-1 and at the quantitation limits, respectively. The methods are selective, sensitive and precise enough for determination of 4-5 ng ml-1 of nitecapone in plasma and urine and are thus suitable for the kind of pharmacokinetic studies exemplified in this paper.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 50 条