Comparison of Automatic and Manual Threshold Testing in Patients with Permanent Pacemakers: Results from COMET Study

被引:0
|
作者
Ando, Kenji [1 ]
Miura, Fumiharu [2 ]
Masani, Fumiaki [3 ]
Kakugawa, Hiroyuki [4 ]
Tejima, Tomoyuki [5 ]
Doi, Kentaro [5 ]
机构
[1] Kokura Mem Hosp, Dept Cardiol, Kitakyushu, Fukuoka, Japan
[2] Hiroshima City Hosp, Dept Cardiol, Hiroshima, Japan
[3] Niigata Prefectural Cent Hosp, Dept Cardiol, Johetsu, Japan
[4] Natl Hosp Org, Hamada Med Ctr, Dept Cardiol, Hamada, Japan
[5] Medtron Japan Co Ltd, Cardiac Rhythm Dis Management, Tokyo, Japan
关键词
Pacemaker; Automaticity; Pacing threshold;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Automaticity of cardiac devices is designed to reduce the burden on physicians while securing patient safety. The aim of this study is to elucidate accuracy of automatic adjustment algorithm of pacing output and effectiveness of this feature for time reduction during clinic visits. Methods and Results: Patients implanted with pacemakers capable of automatically adjusting the pacing output were enrolled. During their routine device follow-ups, pacing thresholds were compared between manual measurements and most recently recorded data in the device, which had been measured automatically at prespecified timing. Difference between -0: 25 and 0.5 volts were predefined as equivalent. Time required for conventional follow-up was also compared to device interrogation and data assessment. A total of 359 patients from 32 centers were enrolled. Calculated 95% confidence interval of difference in volts for atrium was (-0: 0625; 0) and the difference for ventricle was nonsignificant. These results met the equivalence criteria. Observed time to interrogate and assess device-stored pacing threshold was shorter by average of 3.3 minutes compared to manual method. Conclusions: The pacing threshold determined by the automatic measurement algorithm was accurate. Follow-up time was shorter when utilizing automaticity of pacemakers. The system is safe and time efficient.
引用
收藏
页码:307 / 313
页数:7
相关论文
共 17 条
  • [1] Uni- and bipolar pacing threshold measurements with capturecontrol, a new automatic pacemaker function for capture verification - Comparison of the automatic and the manual pacing threshold measurement
    Schuchert, A
    Meinertz, T
    CARDIOLOGY, 1999, 92 (03) : 210 - 213
  • [2] Trends in the use, complications, and costs of permanent pacemakers in Australia: A nationwide study from 2008 to 2017
    Westaway, Samuel
    Nye, Elsbeth
    Gallagher, Celine
    Tu, Samuel J.
    Clarke, Nicholas
    Hanna-Rivero, Nicole
    Emami, Mehrdad
    Kadhim, Kadhim
    Pitman, Bradley M.
    Mahajan, Rajiv
    Lau, Dennis H.
    Young, Glenn D.
    Sanders, Prashanthan
    Wong, Christopher X.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (02): : 266 - 273
  • [3] What is the minimal pacing rate that prevents torsades de pointes?: Insights from patients with permanent pacemakers
    Pinski, SL
    Eguía, LE
    Trohman, RG
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (11): : 1612 - 1615
  • [4] An observational study of the effect of coronary artery disease severity on acute pacing threshold and lead impedance in patients with permanent pacemaker
    Ekinci, Selim
    Yesil, Murat
    Bayata, Serdar
    Arikan, Erdinc
    Vurgun, Veysel Kutay
    Postaci, Nursen
    Safak, Ozgen
    Ozdemir, Emre
    ANADOLU KARDIYOLOJI DERGISI-THE ANATOLIAN JOURNAL OF CARDIOLOGY, 2012, 12 (03): : 208 - 213
  • [5] Interference between dental electrical devices and pacemakers or defibrillators Results from a prospective clinical study
    Elayi, Claude S.
    Lusher, Stephanie
    Nyquist, Jillian L. Meeks
    Darrat, Yousef
    Morales, Gustavo X.
    Miller, Craig S.
    JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2015, 146 (02) : 121 - 128
  • [6] Quality of life and exercise capacity in patients with prolonged PQ interval and dual chamber pacemakers: a randomized comparison of permanent ventricular stimulation vs intrinsic AV conduction
    Maurer, DE
    Naegeli, B
    Straumann, E
    Bertel, O
    Frielingsdorf, J
    EUROPACE, 2003, 5 (04): : 411 - 417
  • [7] Factors Influencing Pacemaker Generator Longevity: Results from the Complete Automatic Pacing Threshold Utilization Recorded in the CAPTURE Trial
    Rosenthal, Lawrence S.
    Mester, Steven
    Rakovec, Peter
    Penaranda, J. Benezet
    Sherman, Jon R.
    Sheldon, Todd J.
    Zeng, Cathy
    Wang, Paul
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (08): : 1020 - 1030
  • [8] Incidence of ventricular tachyarrhythmias during permanent pacemaker therapy in low-risk patients results from the German multicentre EVENTS study
    Faber, Thomas S.
    Gradinger, Robert
    Treusch, Sven
    Morkel, Carsten
    Brachmann, Johannes
    Bode, Christoph
    Zehender, Manfred
    EUROPEAN HEART JOURNAL, 2007, 28 (18) : 2238 - 2242
  • [9] Non-invasive assessment of myocardial contractility from force-frequency relationship in patients with implanted pacemakers: first results
    Bondke, Hansjuergen
    Borges, Adrian-Constantin
    Petersen, Sven
    Walde, Thomas
    Baumann, Gert
    EUROPACE, 2010, 12 (07): : 968 - 971
  • [10] Comparison of Fontan Survivors with and without Pacemakers: A Report from the Pediatric Heart Network Fontan Cross-sectional Study
    Williams, Richard V.
    Travison, Thomas
    Kaltman, Jonathan R.
    Cecchin, Frank
    Colan, Steven D.
    Idriss, Salim F.
    Lu, Minmin
    Margossian, Renee
    Reed, John H.
    Silver, Eric S.
    Stephenson, Elizabeth A.
    Vetter, Victoria L.
    CONGENITAL HEART DISEASE, 2013, 8 (01) : 32 - 39